8 DePuy Synthes 3.5 mm LCP
®
Extra-articular Distal Humerus Plate Surgical Technique
3
Position plate on bone
Instruments
329.02*
Bending Iron, for 4.5 mm plates (2 required)
329.30*
Plate-Bending Press
399.121*
Bone Holding Forceps, 14 mm, serrated jaw,
soft ratchet
Position the plate so that the shaft portion of the plate is
located centrally on the posterior aspect of the bone while
the distal end curves along the back of the lateral column.
Ensure that the plate is at a safe distance from the olecranon
fossa so that complete elbow extension is not impeded.
The position of the plate should allow distal screw insertion
through the lateral flange to reach far into the trochlea.
Due to varying patient anatomy, slight bending may be
necessary. Contour the plate as needed using the bending
irons or the plate-bending press.
Reduce Fracture
1
Reduce fracture and fix temporarily
Use pointed forceps for temporary fixation in restoring
the anatomy. Ensure that forceps will not interfere with
subsequent plate placement.
2
Determine plate length
Choose a plate length that offers sufficient fixation proximal
to the fracture.
* Also available