Page 1: ...ss underutilized and idle equipment along with credit for buybacks and trade ins Custom engineering so your equipment works exactly as you specify Critical and expedited services Leasing Rentals Demos...
Page 2: ...Sonic Meter SM1000 Operating Manual 2 00 1 11 Rev 1 1 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 3: ...ttempt to operate this instrument without completely reading this manual and understanding the correct operating procedures For online technical support and further information refer to our web site a...
Page 4: ...s 4 Operation 4 1 Changing the batteries 4 2 Resetting the circuit breaker 4 3 Powering up the meter 4 4 Options 4 5 Battery capacity 4 6 Real time reading 4 7 Averaged reading 4 8 Setup 4 9 Powering...
Page 5: ...1 4 Probe The probe attaches to one end of the 1 meter co axial BNC cable The other end of the BNC cable that connects to the meter Never attempt to remove or install the probe or cable while the mete...
Page 6: ...an be referred to as hot or cold spots in the liquid Hot and cold spots are locations of high and low intensity scrub force relative to a mean value scrub force These process variations can significan...
Page 7: ...000 3 1 Sonic Meter The Sonic Meter SM1000 is packaged in an industrial T shaped housing The housing is made of hard anodized aluminum The housing is splash proof not waterproof 3 2 Keypad Artisan Tec...
Page 8: ...ice and for scrolling right when changing input values Power OFF Scroll Left Color designation is red with 0 Used for powering off the meter for returning back to the previous menu and for scrolling l...
Page 9: ...itch the probes power up the meter and then go into the Setup menu and change the probes sensitivity This procedure is explained in detail in section 4 8 3 6 Battery charger The battery charger can on...
Page 10: ...d on the bottom of each battery socket Reinstall the battery cover with the retaining tab positioned first in the bottom of the meter before securing the cover If the batteries are installed with the...
Page 11: ...Figure 4 shows the circuit breaker located at the bottom of the meter just below the battery housing Remove the battery cover With a small screwdriver push the small white tab of the circuit breaker...
Page 12: ...atteries drop off quickly With continuous use the meter will run approximately 8 hours 4 6 Real time reading In Figure 5 the Options menu displays the selection arrow at the 1st position Note the 1st...
Page 13: ...right hand corner Refer to Figure 8 This indicator appears when the pressure reading is less than the acceptable High range limit value XXX XXpsi For example If the High range limit is 25 00psi and t...
Page 14: ...the averaging of the pressure based on the number of pressure reading samples and the time interval Selections 2 will allow the user to select the number of samples Selection 3 will allow the user to...
Page 15: ...git press and release the red Scroll Left key once to return the selection arrow to the 2nd position The value will automatically be saved in non volatile memory To set the time interval between sampl...
Page 16: ...roll Right key once The following screen prompt will appear as shown below in Figure 15 This screen will remain for the duration of the averaging The index number of samples of the average will count...
Page 17: ...ach digit can be changed To move to the next digit use the red Left Scroll key The cursor can always be moved to the previous digit by using the green Right Scroll key After the last digit hundreds ha...
Page 18: ...At the Setup menu with the selection indicator at the 3rd position Probe press and release the green Scroll Right key The following median setup screen will appear as shown in below in Figure 22 Figu...
Page 19: ...er of samples to 10 and the interval to 1 second These settings will average 10 samples over a 10second period With the meter still in averaged reading mode take 8 to 10 readings throughout the tank M...
Page 20: ...tervals of 10msec to 99 99sec Accuracy The Sonic Meter SM1000 uses a 12bit ADC to digitize the analog signal coming from the probe s transducer The accuracy of the meter is 0 03psi Battery life hours...
Page 21: ...Craft LLC will at its option repair or replace products that prove to be defective The foregoing warranty shall not apply to defects resulting from improper use or modification No other warranty is ex...
Page 22: ...uipment Have surplus equipment taking up shelf space We ll give it a new home Learn more Visit us at artisantg com for more info on price quotes drivers technical specifications manuals and documentat...