SynCardia TAH-t Instructions for Use
SynCardia Systems, Inc.
Page 21
•
DO NOT
leave key in primary controller key switch during an implant.
The key may be kept on Velcro near top right side of the backup air
supply compartment.
•
DO
have a backup system in a state of ready standby.
•
DO
set backup controller parameters to the same values as the primary
controller.
•
DO
have a controller switch key attached to the console.
•
DO
keep system casters locked except for transport.
•
DO
have a number of spare charged air tanks on hand.
8.2
Readying Console for Clinical Use
•
Two consoles should be in ready standby mode. Ensure that backup
batteries are fully charged.
•
Verify that each system has been connected to AC power with the
SYSTEM POWER switch in the ON (1) position
•
Confirm that each controller AC POWER/BATT CHRG lamp is on.
•
Before moving system to Operating Room, moisten a clean cloth with
an antibacterial agent and wipe down all exterior surfaces of the
console.
•
Do not spray any cleaning agent directly on system.
8.3
SynCardia TAH-t Startup Procedures
•
Turn SYSTEM POWER switch OFF (0); disconnect system mains
power cord and move system to patient site.
•
Position the rear side of the console within driveline length of the
patient’s chest.
•
Lock front casters (wheels).
•
Connect system power cord and turn SYSTEM POWER switch ON
(1).