12
GETTING TO KNOW EZ DAM
®
Sterilization and disinfection of the parts
1.
Disconnect the mouthpiece and discard.
2.
Disconnect the EZ DAM
®
core and bite block from the valve.
3.
Remove the bite block from the EZ DAM
®
core.
4.
Autoclave the EZ DAM
®
core and bite block in separate sterilization
pouches. Do not use dry heat mode. Keep the temperature between
120°C and 132°C. The EZ DAM
®
core can also be cold soaked.
5.
Run disinfectant solution through the suction hose after each use.
Do not autoclave the suction hose.
6.
Store items in sterilized pouches until the next use.
7.
All parts must be completely dry and cooled down before use.
Important, doing any of the following will void warranty:
• DO NOT stack items or place heavy instruments on top of bite blocks or
EZ DAM
®
core during sterilization.
• DO NOT autoclave above 132°C.
• DO NOT chemiclave.