Chapter 2 Performance Tests
58502A Operational Verification
2-6
User’s Guide
3
Connect the equipment as shown in Figure 2-1. (Connect the Signal
Generator output to the 58502A’s input
A
or
B
rear-panel input, and
connect oscilloscope Channel 1 to output
1
port of the 58502A.
Figure 2-1. Operational Verification Setup
4
Verify that the front-panel
Input A
or
B
and all twelve
OUTPUT
STATUS (1-12)
output LEDs are illuminated.
5
Set the oscilloscope to the following:
Channel 1: 0.5 V/div
Timebase: 50 ns/div
6
Connect the oscilloscope to each of the rear-panel output ports in turn
to verify that the output signal is a nominal 1 V rms (2.8 Vp-p) sine
wave at 10 MHz.
7
Disconnect the signal from the 58502A rear-panel input connector.
8
Verify that all input and output LEDs are off.
This indicates loss of input and output.
9
Mark Pass or Fail in Line 1 on the Operational Verification portion of
the 58502A Performance Test Record, located on page 2-14.
HP 54600B
Oscilloscope
or equivalent
58502A
Distribution
Amplifier
50
Ω
Feedthrough
(HP 10100C)
BNC Cable
BNC
Cable
HP 8663A
Synthesized Signal Generator
or equivalent
Output
Outputs
1 - 12
N-to-BNC
Connector
(1250-0780)
Input
A or B
A
B
1
2
3
4
5
6
7
8
9
10
11
12
5 VRMS MAX
100-240 VAC 50/60GHz
40 VA Max
!
!
!
Summary of Contents for 58502A
Page 10: ...In This Guide x User s Guide ...
Page 11: ...1 Getting Started ...
Page 27: ...2 Performance Tests Verifying Specifications ...
Page 29: ...Chapter 2 Performance Tests Introduction User s Guide 2 3 ...
Page 47: ...3 Features and Functions ...
Page 59: ...4 Command Reference ...
Page 73: ...5 Functional Description ...
Page 78: ...Chapter 5 Functional Description Functional Block Description of the 58502A 5 6 User s Guide ...
Page 79: ...6 Service ...
Page 81: ...7 Specifications ...
Page 84: ...Chapter 7 Specifications Introduction 7 4 User s Guide ...