Swiss manufacturer of precision measuring instruments since 1969
2
Restore factory settings
This function allows you to revert the D70 to default settings.
Caution: This function will reset all of the parameters and settings of your D70. Tolerances, benchmark settings
and sensor settings will be lost. (Linearisation is not lost on the D70S version however; it simply needs to be
reactivated.)
Mini-USB connector
This can supply power to the D70 and also allows for the
transmission of measurements to a computer.
Install the drivers located on the CD-ROM which accompa-
nies the D70. Then connect the unit to the computer. The
computer should recognise the new device and will display
the following message:
To do this, follow the procedure below:
1. Turn off the D70.
2. Switch on the D70.
3. When the start screen appears, tap the Sylvac logo.
4. A panel with five icons will appear.
5. Tap Linearity.
6. To activate the linearisation function, choose “En-
abled” = “Yes”.
7. You may then enter up to 25 correction points.
8. To do this, place yourself in a position corresponding to
one point: the non-corrected sensor value will be dis-
placed in the “Probe” box. Input the actual value in the
“Real position” box.
9. Press the “Home” icon to go back
To do this, follow points 1 to 4 above, then press “Initialisation” and confirm with YES or cancel
with NO. Then tap the “Home” icon to return to the measurement screen.
Your devices are ready to use
Device driver software installed
successfully.
681.124-110