Guidelines and Manufacturer’s Declaration: 
Electromagnetic Stability Interference 

The iMRS one is intended for operation in an ELECTROMAGNETIC ENVIRONMENT 
as indicated below. The customer or iMRS user should ensure that it is operated 
in such an environment.

Interference 

Immunity Tests

IEC 60601 

Test Level

Com-

pliance 

Level

Electromagnetic Environment - Guidelines

Conducted RF 

interference 

acc. to IEC 

610004-4-6

Radiated RF 

interference 

acc. to IEC 

610004-4-3

3 V effec-

tive value

150 kHz up 

to 80 MHz 

3 V/m 

80 MHz up 

to 2.5 GHz

3V

Portable and mobile radio devices, includ-

ing their cables, should not be used at 

a distance closer to the iMRS one than 

recommended, which has been calculated 

according to relevant equation for the trans-

mission frequency.

Recommended safe distance:

                   for 80 MHz up to 800 MHz

                   for 800 MHz up to 2.5 GHz

Where P is the rated power of the transmit-

ter in watts (W) according to the information 

from the transmitter manufacturer and d is 

the recommended safe distance in meters 

(m). The field strength of stationary radio 

transmitters should be investigated locally 

for all frequencies lower than the compli-

ance level 

6

Interference is possible in proximity  

to devices that bear the following  

symbol.

NOTE 1 The higher frequency range applies at 80 MHz and 800 MHz.

NOTE 2 These guidelines may not apply in all cases. The expansion of electromagnetic 

quantities will be affected by absorption and reflection from buildings, objects and 

people.
The field strength of stationary transmitters, such as: the base stations of cordless 

telephones and land mobile radio systems, amateur radio stations, AM/FM radio and 

television transmitters; cannot be determined in advance with theoretic precision. A 

study of the electromagnetic phenomena of the location should be considered in order 

to determine the nature of the ELECTROMAGNETIC SURROUNDINGS in terms of sta-

tionary transmitters. If the field strength measured at the location where the iMRS one 

will be used exceeds the COMPLIANCE LEVEL mentioned above, the iMRS one should 

be checked to verify its OPERATION in the manner intended. If unusual performance 

characteristics are observed, additional measures may necessary, such as changing the 

orientation or choosing a different location for the iMRS one.

The field strength should be less than 3V/m over the frequency range 150 kHz to 80 

MHz.

6)

 National footnote: User here is meant in the sense of RESPONSIBLE ORGANISATION.

d = 1,2√P

d = 1,2√P

d = 2,3√P

Summary of Contents for Omnium1 iMRS one 2.0

Page 1: ...intelligent Lifestyle Operating Manual 2 0...

Page 2: ...the system when sold Copyright Copyright 2019 Swiss Bionic Solutions Holding GmbH All rights reserved No part of this manual including the products and software described herein may be reproduced tran...

Page 3: ...13 6 Activation 14 7 Saving and Loading Pre set Parameters 15 7 1 Saving 15 7 2 Loading 16 8 Starting an Application 17 9 Quick Start Programs 18 10 Settings 18 10 1 About 18 10 2 Factory Settings 19...

Page 4: ...lds and the frequencies in the ionosphere Schumann reso nances exclude all possible danger Note about electromagnetic tolerance EMC Medical electric devices are subject to special precautions in terms...

Page 5: ...ration in an ELECTROMAGNETIC ENVIRONMENT as shown below The customer or user of the iMRS one should ensure that it is operated in such an environment Interference emission mea surements Compliance Ele...

Page 6: ...or Not applicable The quality of the supply voltage should correspond to that in typical business or hospital surroundings Voltage dips short inter ruptions and fluctuations in the supply voltage acc...

Page 7: ...ance level 6 Interference is possible in proximity to devices that bear the following symbol NOTE 1 The higher frequency range applies at 80 MHz and 800 MHz NOTE 2 These guidelines may not apply in a...

Page 8: ...ansmitter W Safe distance m dependent on transmitter frequency 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 F...

Page 9: ...healing reaction arises in approximately 10 of the patients treated in the first days or weeks of application such as through an increase in the symptoms This should be frequently expected after prol...

Page 10: ...e and should therefore be avoided 3 Possible Applications The iMRS one magnetic resonance stimulation system can be used For wellness purposes and to generally increase vitality and well being For pal...

Page 11: ...e damaged or missing please contact your consul tant at Swiss Bionic Solutions iMRS one Android software Pre installed on the Omnium1 OmniMat OmniPad OmniSpot Not included with the iMRS one basic set...

Page 12: ...ol unit The Omnium1 control device is included in the scope of delivery for all iMRS one sets available You will find the operating instructions for the Omnium1 in the special manual that accompanies...

Page 13: ...D A converter with the 20 pin ribbon cable Connect the second plug attached to the 20 pin ribbon cable with the corresponding connection on the Omnium1 see the Omnium1 operating manual Connect the de...

Page 14: ...ate activate the iMRS one app by tapping the program icon 1 on the Omnium1 desktop 2 Afterwards the iMRS one initial screen will display the legally indicated contraindications As soon as it has been...

Page 15: ...rtual keyboard will disappear You can now start making settings for the first applicator with the first organ clock setting To do so tap the circle next to Timer 3 and set the desired application dura...

Page 16: ...the users desired you can tap Back 1 to return to the main screen for the iMRS one application 7 2 Loading Choose the user once the system has successfully been started To do that tap on the User icon...

Page 17: ...rwards tap the circle next to Intensity 4 and set the desired inten sity of the magnetic field flux density This setting has been divided into the levels Sensitive 10 25 50 100 150 200 and 400 The int...

Page 18: ...ield application will start immediately after you have touched one of the five quick start buttons All of the necessary parame ters have already been defined 10 Settings Tapping on the Settings button...

Page 19: ...of windings intensity have been built into the full body applicator The copper coil pair at the head cable connected to the applicator box has the lowest number of windings and thereby generates the l...

Page 20: ...irculation of air must be ensured as a pro tective measure against the formation of moisture The magnetic field not only works directly above the pillow applicator but also spreads itself out in all d...

Page 21: ...creen You can remove most remnants with a circular motion For rough dirt lightly moisten the cloth and clean the affected area with that Note Be very careful that water does not enter the housing beca...

Page 22: ...een connected properly Error No DA Converter Connected The D A converter has not been connected or not been connected properly Goggle does not connect The OMNIBRAIN system has not been connected or no...

Page 23: ...6299 SN MH2005 00804 0483 MRS 2000 designo Home r MediConsult GmbH Am Sandfeld 17a 76149 Karlsruhe Deutschland ng 12 V 50 Hz 8 VA nur mit Original MediConsult Zubeh r und netzteil FW6299 2005 00804 04...

Page 24: ...27 mm 21 Temperature in use 10 C bis 40 C 22 Temperature stored 20 C bis 45 C 23 Air humidity in use 30 75 RH without condensation 24 Air humidity stored 10 95 RH without condensation 25 Air pressure...

Page 25: ...65 00 120 00 16 Warranty The iMRS one magnetic resonance stimulation system is the result of innovative research and development work Swiss Bionic Solutions pro vides the warranty on the applicators a...

Page 26: ...26...

Page 27: ...27...

Page 28: ...de swissbionic com Swiss Bionic Solutions USA Inc 12330 SW 53rd Street Suite 703 704 Cooper City Florida 33330 USA Phone 1 954 766 4153 Fax 1 954 766 4156 Email us swissbionic com Swiss Bionic Soluti...

Reviews: