Poor image quality 

No cover slip on the 
specimen 

Place glass cover slip over 
the specimen 

 

Slide is too thick or too thin  Use a lab-quality slide 

 

Slide is placed upside 
down on the stage 

Flip the slide over so that 
the cover slip is facing up 

 

Oil on objective lens 

Wipe the lens off with soft 
cloth and glass cleaner 

 

Iris aperture is too narrow 
or too wide 

Adjust the iris aperture 

Image moves while 
adjusting focus 

Slide is shifting on the 
stage 

Reposition slide more 
securely in the slide holder  

Light is too dim 

Iris aperture is closed too 
far 

Open the iris aperture 
wider 

 

Bulb setting is too low 

Increase intensity on the 
control wheel 

 

Condenser lens is dirty 

Wipe the condenser lens 
off with a soft cloth 

 

Voltage is too low 

Use the supplied power 
adapter 

Objective lens collides with 
the slide when switching 
from a lower to a higher 
power objective 

Slide is upside down or too 
thick 

Use a lab-quality slide or 
flip the slide over so the 
cover slip is facing up 

 

Stage is set too close to 
the objectives 

Lower the stage with the 
coarse focus knob 

 

Safety rack stop needs 
adjustment 

Adjust the safety rack stop 
to stop the stage farther 
from the objectives 

Image won't come into 
focus when using a 
higher-powered objective 

Slide is upside down 

Flip the slide over so the 
cover slip is facing up 

 

Summary of Contents for SW380B

Page 1: ...User Guide SW380B...

Page 2: ...e components of your new microscope and help you begin to use it For more information about our microscopes or customer service requests please visit www swiftoptical com Table of Contents I Specifica...

Page 3: ...65 working distance 0 63mm 100X Spring Oil NA 1 25 working distance 0 195mm Magnifications 40X 100X 250X 400X 1000X and 2500X Stage Double layered mechanical stage with slide holder Stage size 130mmx...

Page 4: ...ng care not to overtighten Remove the plastic covers from the eyepiece holders on the microscope s head then place eyepieces into the holders Locate the power cord and attach it to the microscope then...

Page 5: ...lder once the specimen is in focus Increase magnification by rotating the nosepiece clockwise to a higher magnification objective taking care not to strike the slide with the objective as doing so may...

Page 6: ...disassemble the electrical mechanical or optical components This servicing should only be done by a Swift technician Issue Possible cause Solution Visual field is uneven or image is dark at the edges...

Page 7: ...n the iris aperture wider Bulb setting is too low Increase intensity on the control wheel Condenser lens is dirty Wipe the condenser lens off with a soft cloth Voltage is too low Use the supplied powe...

Page 8: ...r Microscopes Microscopes come with a five 5 year warranty against manufacturing defects Does not cover normal wear routine maintenance add on accessories damage resulting from repair by unauthorized...

Page 9: ...tment Tel 877 967 9438 option 1 Email customersupport swiftoptical com Disclaimer We are constantly working to improve our instruments and to adapt them in response to customer feedback These improvem...

Reviews: