background image

 

COMPONENTS OF THE STEREO MICROSCOPE 

 

BOTTOM ILLUMINATOR – The light source that is mounted in the 

base to provide trans-illumination. 

 

DIOPTER ADJUSTMENT – Located on the left eyepiece tube (SM90) 

or left and right eye tubes (M28Z/M29TZ).  This adjustment is used 

to fine focus the optics to compensate for visual differences 

between the user’s eyes. 

 

EYEPIECE – The upper optical element that further magnifies the 

primary image of the specimen and brings the light rays in focus at 

the eyepoint. 

 

EYEPIECE TUBE – The part of the head that holds the eyepieces in 

place.  Set screws in the eyepiece tubes can be used to lock the 

eyepieces in place.  

 

STAGE CLIP – Used to hold specimen in place. 

 

FOCUSING CONTROL KNOB – The focusing control knobs located on 

either side of the stand are used to raise or lower the head to bring 

the specimen into focus.  The control knobs incorporate a slip 

clutch system which prevents the head from traveling past both 

ends of the focusing range.  When the focusing limit is reached, the 

clutch system begins to slip to prevent gear damage. 

 

HEAD – The 360˚ rotatable component that contains all of the 

optical elements of the stereo microscope including the eyepieces, 

refracting prisms and objective lenses.  The head of the SM90 series 

and M28Z / M29TZ is designed to allow users to adjust the 

interpupillary distance of the eyepieces for proper viewing. 

 

ILLUMINATOR DIMMER CONTROL – The intensity of the top 

(incident light) LED or bottom (trans-illumination) LED can be 

adjusted to the preferred brightness (0 through 9) by using the 

dimmer control located on the side of the base. 

 

ILLUMINATOR ON/OFF SWITCHES – A tri-illumination system is 

incorporated into the stand.  The illumination system is controlled 

by top (incident light) and bottom (trans-illumination) on/off 

switches located at the base of the stand.  The incident illuminator 

is used for opaque specimens, while the trans-illuminator 

effectively illuminates the internal structure of transparent 

specimens.  Translucent specimens may be seen in greater detail if 

both illuminators are used simultaneously.   

Summary of Contents for M28Z

Page 1: ...Swift SM90 M28Z M29TZ Series Stereo Microscope Use and Care Manual SWIFT OPTICAL Enduring Quality and Technical Excellence...

Page 2: ...um versatility of the instrument for classroom use It is built to the highest most rigid optical and mechanical standards SWIFT M28Z M29TZ ZOOM STEREO MICROSCOPES The Swift M28Z M29TZ zoom stereo micr...

Page 3: ...ing Distance 80mm 80mm 80mm Stand SM90CL SM90CL SM90CL SM90CL Illuminator incident transbase 3W 5500K LED 3W 5500K LED 3W 5500K LED M28Z M29TZ STEREO ZOOM MICROSCOPE SPECIFICATIONS Model Number M28Z M...

Page 4: ...d Top incident Illuminator Focusing Control Knob Illuminator On Off Switches Illuminator Dimmer Control Turret Stage Clip Stage Plate Stand SM91 SM90CL Pictured Diopter Adjustment Bottom transmitted I...

Page 5: ...5 M28Z SM90CL pictured Magnification Knob Diopter Adjustment...

Page 6: ...h ends of the focusing range When the focusing limit is reached the clutch system begins to slip to prevent gear damage HEAD The 360 rotatable component that contains all of the optical elements of th...

Page 7: ...ES STEREO MICROSCOPE 1 Place the specimen onto the stage plate and select the type of illumination If the specimen is transparent turn on the bottom illumination If the specimen is opaque turn on the...

Page 8: ...ine on the eyepiece tubes 3 Grasp the eyepiece tubes and move them either closer together or farther apart to see one clear image Note if two separate images are observed the eyepiece tubes are too fa...

Page 9: ...top bulb may be replaced by carefully laying your microscope on its back so that the top light is facing up Unscrew the top light cover to expose the LED bulb assembly Carefully remove the LED conden...

Page 10: ...lly with quality lens tissue If dirt or other foreign matter still remains it may be necessary to use water or optical cleaner eyeglass or camera lens Note the lens tissue should be moistened not satu...

Page 11: ...e positioned to be in the optical path C PROBLEM Image of the specimen goes out of the focus all by itself CORRECTION Use the tension control collar to tighten the focusing mechanism found on the focu...

Page 12: ...AL INSTRUMENTS INC LIMITED LIFETIME WARRANTY Please see our website www swiftoptical com for complete warranty details and exclusions Swift Optical Instruments Inc 877 967 9438 www swiftoptical com Re...

Reviews: