AMU pH-Redox
Program List and Explanations
The point of intersection of the tangent with the respective axis will
result in the parameters a and L.
Consult the manual of the control unit for connecting and program-
ming details. Choose control upwards or downwards.
If Control upwards or Control downwards is active
5.2.1.43
Control Parameters
5.2.1.43.10
Setpoint:
User-defined process value (Measured value or flow)
5.2.1.43.20
P-Band:
Range below (upwards control) or above (downwards con-
trol) the set-point, within the dosing intensity is reduced from 100%
to 0% to reach the set-point without overshooting.
5.2.1.43
Control Parameters:
if Parameters = Meas. Value
5.2.1.43
Control Parameters:
if Parameters = Temperature
5.2.1.43.11
Setpoint
: -25 °C to +270 °C
5.2.1.43.21
P-Band:
0 °C to +100 °C
5.2.1.43
Control Parameters:
if Parameters = Sample flow
5.2.1.43.12
Setpoint
: 0.0 l/h –200 l/h
5.2.1.43.22
P-Band:
0.0 l/h –200 l/h
A
B
X
Response to maximum control output
Tangent on the inflection point
Time
Xp
Tn
Tv
= 1.2/a
= 2L
= L/2
X
Y
B
A
L
a
pH sensor
Redox sensor
5.2.1.43.10
5.2.1.43.20
Setpoint
:
P-Band
:
-3 pH to +15 pH
0.00 pH to +2.00 pH
-500 mV to +1500 mV
0 mV to +200 mV