AMU pH-Redox
Error List
A-96.250.441 / 060619
53
7.
Error List
Error
Non-fatal Error. Indicates an alarm if a programmed value is ex-
ceeded.
Such Errors are marked
E0xx
(bold and black).
Fatal Error
(blinking symbol)
Control of dosing devices is interrupted.
The indicated measured values are possibly incorrect.
Fatal Errors are divided in the following two categories:
Errors which disappear if correct measuring conditions are re-
covered (i.e. Sample Flow low).
Such Errors are marked
E0xx
(bold and orange)
Errors which indicate a hardware failure of the instrument.
Such Errors are marked
E0xx
(
bold and red
)
Error or
fatal Error
Error not yet acknowledged.
Check
Pending Errors 1.1.5 *
and
take corrective action.
Press [ENTER].
Navigate to menu Messages.
Press [ENTER].
Navigate to menu Pending Errors.
Press [ENTER].
Press [ENTER] to acknowledge the
Pending Errors. The Error is reset and
saved in the Message List.
21.7 °C
HOLD
14:10:45
R1
7.76 pH
R2
1
Installation
Operation
Diagnostics
Messages
Maintenance
Main Menu
1.1
Message List
Pending Errors
Messages
1.1.5
Pending Errors
Error Code
E002
Alarm low
<Enter> to Acknowledge