16
A-96.250.131 / 050220
AMI Turbitrace
Installation
3.3. Connect Sample and Waste
3.3.1
Connect Sample Inlet
Use plastic tube (FEP, PA, or PE 4x6 mm) to connect the sample
line.
Mounting of
SERTO fitting
Push the flexible tube [A] into the Serto fitting [D] and tighten the
knurled nut [B].
3.3.2
Connect Sample Outlet
Connect the 1/2” tube [E] to the hose nozzle [D] and place it into a
pressure free drain with sufficient capacity.
A
B
C
D
Serto fitting
Compression ferrule
Knurled nut
Flexible tube
B
C
D
A
A
B
C
D
E
Tube from drain valve
Sample outlet
Drain funnel
Hose nozzle
1/2” tube
B
E
A
C
D