A-96.250.211 / 171121
AMI Sodium P
Instrument Setup
53
4.4. Switch on Power
Let the instrument run continuously for 1 hour with sample.
4.5. Programming
External
devices
Program all parameters for external devices (interface, recorders,
Limits, alarms
Program all parameters for instrument operation (limits, alarms). See
.
2
nd
sample
stream option
If the 2
nd
sample stream option is installed, select the the channel
switching mode. See
for details.
Automatic
regeneration
option
If the automatic regeneration option is installed, configure the time
and days of the week on which an automatic regeneration should be
performed. See
4.6. Fill Tube of Automatic Regeneration Module
If the automatic regeneration module is installed, fill the tube with
regeneration solution by calling the <Maintenance>/<Regeneration>/
<Manual Operation> menu item.
4.7. Perform a Calibration
1
.
2
Calibrate the pH sensor, see
.
3
Etch and calibrate the sodium electrode, see
Standard Sodium 1-Point Calibration, p. 77
.