3.7. Installing the Automatic Regeneration Option
Mechanical
installation
1
Stop the sample flow at the main tap.
2
Switch off the instrument.
3
Remove the elbow piece from the inlet of the flow cell and screw
in the enclosed T-piece [A].
4
Connect the adapter [B] to the T-piece [A].
5
Connect the ventilation tube [D] to the bottle holder of the auto
-
matic regeneration module.
6
Screw the automatic regeneration module to the panel using the
supplied screws (if necessary, unscrew the air filter [C] and screw
it back on at the attachment points further to the left).
7
Use tube 4 to connect the oulet of the bottle holder with the inlet
[E] of the 3-way valve.
8
Use tube 3 to connect the outlet [F] of the 3-way valve with the
inlet [G] of the flow cell.
Note:
The unused port of the 3-way valve must be sealed with a
blind plug.