34
A-96.250.531 / 300617
AMI Oxytrace QED
Installation
3.9.1
Signal Output 3
Terminal 38 (+) and 37 (-).
Requires the additional board for the third signal output 0/4–20 mA
PCB. The third signal output can be operated as a current source
or as a current sink (switchable via switch [A]). For detailed infor-
mation see the corresponding installation instruction.
NOTICE:
Max. burden 510
.
Third signal output 0/4 - 20 mA PCB
3.9.2
Profibus, Modbus Interface
Terminal 37 PB, Terminal 38 PA
To connect several instruments by means of a network or to config-
ure a PROFIBUS DP connection, consult the PROFIBUS manual.
Use appropriate network cable.
NOTICE:
The switch must be ON, if only one instrument is
installed, or on the last instrument in the bus.
Profibus, Modbus Interface PCB (RS 485)
A Operating mode selector switch
A
A On - OFF switch
ON
OFF
A