A-96.250.491 / 190121
AMI Deltacon DG
Maintenance
39
6.4. Replace the Ion Exchanger
The resin of the ion exchanger changes its color from dark violet to
brown if the capacity is exhausted. The resin should be changed be
-
fore no violet resin is left or the cation conductivity rises above the
normal value. At a concentration of 1 ppm alkalization agent, one
resin filling will last for roughly 4 months if sample flow is 10 l/h, or 5
months if sample flow is 5 l/h.
1
Stop sample flow.
2
Slightly squeeze the exhausted cation exchanger bottle [F] be
-
fore removing.
Thus no water will spill out of the flow cell when loosening the
bottle.
3
Unscrew and carefully remove the exhausted cation exchanger
bottle [F].
4
Fill high purity water into the new cation exchanger bottle [D], un
-
til the water level in the bottle reaches the beginning of the
thread.
5
Carefully, without spilling water, push the cation exchanger bottle
over the inlet filter holder [E] into the bottle holder [B].
6
Screw the cation exchanger bottle into the bottle holder.
Do not tighten the bottle too firmly, this could damage the
gasket.
7
In the menu <maintenance/change of resin> set change of resin
to yes.
8
Open the manual deaeration valve [C].
A
B
C
D
E
F
Flow cell
Bottle holder
Manual deaeration valve
New cation exchanger
bottle
Inlet filter holder
Exhausted cation
exchanger bottle
A
B
D
C
E
F