A-96.250.491 / 190121
AMI Deltacon DG
Installation
26
3.7.2
Relay 1 and 2
Note:
Max. load 1 A / 250 VAC
Relay 1 and 2 can be configured as normally open or as normally
closed. Standard for both relays is normally open. To configure a Re
-
lay as normally closed, set the jumper in the upper position.
Note:
Some error codes and the instrument status may influence
the status of the relays described below.
For programming see Menu Installation
Relay
config. Terminals
Jumper
pos.
Description
Relay configuration
Normally
Open
6/7: Relay 1
8/9: Relay 2
Inactive (opened) during
normal operation and
loss of power.
Active (closed) when a
programmed function is
executed.
Normally
Closed
6/7: Relay 1
8/9: Relay 2
Inactive (closed) during
normal operation and
loss of power.
Active (opened) when a
programmed function is
executed.
6
0V
7
6
0V
7
A
B
Jumper set as normally open (standard setting)
Jumper set as normally closed
A
B