FAULT SEARCHING SCENARIO
After cable selecting you can start a fault searching process. This process consists of several
measurements, described in protocol. To
fi
nd a fault you should
fi
ll a protocol.
1. Back wire searching. Connect a
С
wire to cable shield. During open-circuit
fi
nd a wire
with low insulation by the B wire. Insulation resistance is controlled in B-C line. After,
select a good wire by A wire. The insulation resistance is controlled in A-C line.