98
SV 200A User Manual
If different daylight-saving time rules should be applied,
select
Custom
and fill in
Custom Daylight Saving Time
section.
Note: There are four time-synchronization servers predefined in Time synchronization
settings. Do not change these settings unless the servers are unavailable in the
instrument’s network or a custom server is needed
.
9.8.11 Calibration & System Check
SV 200A
is fitted with an
Automatic calibration
and
Automatic system check
features. The
Calibration & System Check
section enables configuring these features.
Automatic calibration
When the
Automatic calibration
is enabled, mounting
on the instrument’s microphone and switching
on the sound calibrator automatically starts the calibration procedure.
Such operation requires typing the
Calibration Level
from the calibration certificate of the used
calibrator. For more information about the automatic calibration
feature see Chapter
Note: Defined Calibration Level remains constant. Remember to perform calibration with
the sound calibrator generating the same level as the configured calibration level.
Automatic calibration
settings
Automatic system
check settings