background image

 

 

 

 

 

USER MANUAL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SV 111 

VIBRATION 

CALIBRATOR 

 

 

Warsaw, 2022-04-21 

Rev. 01.03 

 

Copyright © 20

22 SVANTEK. 

All rights reserved. 

Summary of Contents for SV 111

Page 1: ...USER MANUAL SV 111 VIBRATION CALIBRATOR Warsaw 2022 04 21 Rev 01 03 Copyright 2022 SVANTEK All rights reserved ...

Page 2: ...emarks Trademarks or registered marks in this manual belong to their respective manufacturers Microsoft and Windows are registered trademarks of Microsoft Corporation Disclaimer Information in this document is subject to change without notice and does not represent a commitment on the part of Svantek Svantek provides this document as is without warranty of any kind either expressed or implied incl...

Page 3: ...8 8 Control panel 9 9 Operating diagram 10 10 Submenu diagram 11 11 Calibration of the whole body sensor 12 12 Calibration of the SV 105 Hand Arm sensor 14 13 Calibration of the SV 3023M2 accelerometer 16 14 Calibration of the general purpose accelerometer 17 15 Programming 18 16 General Care and Cleaning 20 17 Technical data 21 18 Declaration of Conformity 22 Appendix A Recalibration 23 Appendix ...

Page 4: ...evice is not in use it is recommended to charge the battery once a month to keep it in good condition It is recommended to carry out a technical inspection and recalibration of the device every 12 months to ensure accurate calibration level Recalibration can be made by the calibration laboratory If there is no possibility to recalibrate the device by calibration laboratory the device should be sen...

Page 5: ... to a certain acceptable level Maximum absolute value of the error of generated vibration signals is called the tolerance and is strictly defined by the standard ISO8041 2005 4 SV 111 model information The model SV 111 is a Portable Vibration Calibrator It is designed to check calibrate various types of vibration level meters according to the ISO 8041 1 2017 Three standard frequencies 15 92 Hz 79 ...

Page 6: ...ve the service staff or the manufacturer directly The complete set includes the following items SV 111 Vibration calibrator SA 33 AC DC adapter Before using the device Be sure that the shaker shipping lock is detached Starting the vibration calibrator with the shipping lock may destroy it To detach the shaker shipping lock unscrew four bolts with the hex wrench and two screwdriver tips attached to...

Page 7: ...d to leave it for a few hours in the room temperature before connecting it to the power supply If any steam condensation appears it is recommended not to connect the device to the mains electricity for 4 to 8 hours until the exterior surface is dry 6 Optional accessories The SV 111 calibrator set doesn t include the SA 105 adapter used for calibration of the SV 105B hand arm transducers and the SA...

Page 8: ...bration calibrator 1 Magnet with two screwdriver tips 2 Adapter 3 Hex wrench slot with support sleeves 4 Charger slot 5 Shaker 6 Transportation lock 7 Support 8 Fastening screw 9 Display 10 Keyboard 11 EXT_DC supply socket 12 USB port 1 2 3 4 5 7 8 9 10 11 12 6 ...

Page 9: ...l 8 Control panel SV 111 control panel 1 Vibrations level button Power on button A 2 Start Stop button 3 Serial number box 4 Display box 5 Vibration frequency button Power button B 6 Spirit level 7 USB port 1 2 3 4 5 6 7 ...

Page 10: ...parameters setting menu Shaker vibration s SV 111 switching on Sensor installation Skipping the levelling process SV 111 levelled properly Sample vibration parameters Start shaker Stop shaker Vibrations parameters not reached properly Shaker vibrates with set parameters ...

Page 11: ...11 SV 111 User Manual 10 Submenu diagram Program submenu can be reached from Vibrations parameters setting menu by pressing To return to Vibrations parameters setting menu press ...

Page 12: ...er from the case cover and screw it to the shaker with one or more bolts one central bolt should be sufficient for the system check dismantle the support system from the case holder put it on the adapter and tight the fixing bolt unscrew the central screw of the whole body sensor and install the sensor on the support system screwing them altogether ...

Page 13: ...User Manual position the sensor according to the calibrated axis channel and tight the fixing bolt put the support system with the tested sensor in the horizontal position on the adapter and tight the fixing bolt ...

Page 14: ...als To perform calibration of the SV 105D Hand Arm sensor use the SA 150C belt with the greatest curvature Screw the SA 150C belt to the SV 105D sensor with the special screw from the SV 105D kit using for that a special 1 5mm Allen screwdriver both included in the SV 105D kit If the belt was screwed to the sensor with the normal screw it is necessary to unscrew it and screw again with the SA 105D...

Page 15: ...on sensor to the calibrator s shaker using special stud included in the SV 105 kit Positioning of the adapter with the sensor for calibration of the X axis Y axis and Z axis should be as per photos below For this use the appropriate hole in the adapter for the stud X Y Z ...

Page 16: ...er with special adapter SA 44 The SA 44 adapter set includes two studs M5 M5 and M5 M3 The M5 M3 stud is used for fixing the accelerometer to the adapter Another M5 M5 stud is used for fixing adapter with the accelerometer to the shaker The SA 44 adapter allows to calibrate the SV 3023M2 accelerometer in all directions ...

Page 17: ... User Manual 14 Calibration of the general purpose accelerometer The general purpose accelerometer for example SV 80 is mounted directly on the shaker with special stud normally included in the accelerometer set ...

Page 18: ...switches to vibrations parameters setting menu It is possible to skip leveling process but it is strongly recommended to proceed the leveling process before every calibration Spirit level can be launched anytime using the program submenu After leveling the device displays the default vibrations parameters values You can change values by pressing the Freq button in the following order 15 92 79 58 1...

Page 19: ...ions are set with 16 Hz 1 m s2 120 dB then the total vibrations amount with frequencies n 16 HZ 32 48 64 80 cannot be higher than 0 95 m s2 94 dB SV 111 is compensating harmonic distortion X Y Vibration means that vibrations level in X or Y corresponding diode is lighting red direction are higher than 10 20dB of vibrations level generated in Z direction This means that both kinds of mistakes were ...

Page 20: ...ect the device from the power supply Wipe the device s surface with the cloth damped with the mixture of warm water and detergent After cleaning wipe the device with dry cloth and wait until the surface is completely dry Do not immerse the device in any fluids as this may damage the device and cause electrical shock Only the external parts of the device should be cleaned ...

Page 21: ... main direction Harmonic distortion 5 3 3 3 General Maximum weight of tested object 1000 300 200 200 g Sensor mounting Threaded hole M5 x 12 mm Seat adapter for SV 100 SV 38 SV 38V Mounting disc for attaching with Beeswax or SA 38 adapter Levelling off time Typically 15 20 seconds 60 seconds max Working conditions Temperature range 10 C 50 C Humidity range 25 85 Power supply Battery type Rechargea...

Page 22: ...22 SV 111 User Manual 18 Declaration of Conformity ...

Page 23: ...15 92 Hz Install the reference sensor and start the shaker with 15 92 Hz and 1m s2 vibrations parameters Wait until the vibrations are steady diodes are lighting with green continuous light and information Level OK is displayed Read the vibrations level value from the reference sensor When it is the same as the set one skip to the next frequency When it differs it is necessary to enter modify the ...

Page 24: ...enu positions displayed are o Leveling o Calibration coefficient that was previously entered o Position where the new calibration coefficient can be set or Checking currently entered calibration coefficient o when the CALIBR VIEW is displayed o press the Level and Freq buttons at the same time o Currently entered value will be displayed on the screen Changing the calibration coefficient o When a n...

Page 25: ...ue by 0 05 dB press the Freq button o Currently set value will be displayed on the screen o To confirm the set value press the Level and Freq at the same time o To save the set value press Level o To exit without saving press Freq After saving confirming screen will be displayed As the calibration coefficient is set it is recommended to repeat the measurement with reference sensor ...

Page 26: ...tion coefficient Use one of formulas bellow 𝑪 𝟐𝟎 𝐥𝐨𝐠𝟏𝟎 𝑨 𝑨𝟎 dB Where A reference sensor vibration level A0 selected vibration level of the calibrator C Ar Ac Where Ac selected vibration level of the calibrator dB Ar reference sensor vibration level dB ...