background image

 

10 

 

Figure 2-1. Liquid light guide installation. 

The output lens functions to recollimate the beam. By correctly positioning the light 
guide sleeve in the output lens tube, the light guide output can be collimated (focused at 
infinity). This is most easily done by shining the light (preferably filtered to remove UV) 
at a uniform surface at least ten feet away and adjusting the position of the end of the 
light guide to a point where the light casts a well-focused circle on the distant surface. At 
10 feet, this circle will be about 6 inches in diameter, have sharp edges, be uniform, and 

be quite brilliant if unfiltered.  
The output lens tube is installed into one or more adapter pieces that mechanically and 
optically couple the light into the microscope. In external adapters, the adapter system 
mounts in place of the manufacturer’s lamp house typically on the back of the 
microscope. In internal adapters, the adapter system is integral to the microscope, and 
the output lens is positioned inside the microscope as close as possible to the back of the 
filter cube-dichroic mirror assembly. This arrangement makes the most efficient use of 

the collimated output beam.  
Refer to documentation included with the individual microscope adapter for further 

installation instructions particular to the adapter. 

CAUTION:  PLEASE NOTE THAT THE LIGHT GUIDE HAS A MINIMUM-

BENDING RADIUS OF 40 MM (ABOUT 1.6 INCHES). 

4.

 

The Analog-IN and/or TTL-IN connections can be made to an external signaling source at 

stage if desired. 

Once all connections have been verified, the system is ready to be turned on.  
 
 

LAMBDA HPX-L5 OPERATION MANUAL – REV. 1.03  (20161130) 

Summary of Contents for Lambda HPX-L5

Page 1: ...igh Power LED Light Source System Air Cooled with Liquid Light Guide Operation Manual Rev 1 03 20161130 One Digital Drive Novato CA 94949 Voice 415 883 0128 Web www sutter com Fax 415 883 0572 Email i...

Page 2: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 2 Copyright 2016 Sutter Instrument Company All Rights Reserved...

Page 3: ......

Page 4: ......

Page 5: ...uppression is recommended at all times Fuse Replacement Replace only with the same type and rating 2 Amp 250V 5 x 20mm Time Delay fuse EIC 60127 2 Example Bussmann GDC 2A or S506 2A RoHS A spare fuse...

Page 6: ...re is any obvious damage to any part of the instrument Avoiding Physical Injury while Powered up and Emitting Light WARNING DO NOT LOOK DIRECTLY INTO THE OUTPUT OF THE LIGHT APERTURE OF THE HIGH POWER...

Page 7: ...ED On Off State while Transitioning Between LED Switch Positions ON and TTL Switch 3 12 2 3 4 Firmware and Lambda HPX Model Switch 4 13 3 OPERATIONS 15 3 1 Indicators 15 3 1 1 Status Display 15 3 2 Co...

Page 8: ...Liquid light guide installation 10 Figure 2 2 Lambda HPX series left side panel removed for access to configuration DIP switches 11 Figure 2 3 Configuration DIP switch bank as viewed head on and top d...

Page 9: ...onnectors The Lambda HPX L5 is liquid light guide coupled version of the Lambda HPX high output LED light source illuminator The liquid light guide allows for LED cooling via an internal heatsink with...

Page 10: ...he MPX L5 Integral shuttering time is 10 microseconds to turn on and off TTL input and output allows for triggering from either software or directly from another device such as a camera or digital I O...

Page 11: ...s a black end sleeve preinstalled over the brass ferrule on each end The two ends of the light guide are identical The end sleeves are designed to slide into the mounting flanges on the input and outp...

Page 12: ...couple the light into the microscope In external adapters the adapter system mounts in place of the manufacturer s lamp house typically on the back of the microscope In internal adapters the adapter...

Page 13: ...ing the first three switches of the four switch DIP bank located on the control board inside the controller cabinet To access the DIP switch bank remove left side panel left side while viewing the fro...

Page 14: ...Switch 2 determines how the state at TTL IN determines the on off state of the lamp when the LED Control Switch is in the TTL position When ON TTL IN going HIGH turns the LED on TTL IN going LOW turn...

Page 15: ...p on ON Lamp turns on Notes 1 and 2 3 Lamp state when switching from manual control to TTL OFF Lamp turns off Note 1 ON Lambda HPX L5 4 Lambda HPX model specific operation Note 3 OFF Lambda HPX Defaul...

Page 16: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 14 This page intentionally blank...

Page 17: ...l indicators controls and connectors on the Lambda HPX L5 controller 3 1 Indicators 3 1 1 Status Display Displays intensity as a percentage and current level in terms of amperes Intensity 5 Current 18...

Page 18: ...s knob dims the LED via Pulse Width Modulation In essence the LED is being switched at 30KHz This is the preferable mode of dimming for most applications as the LED will run cooler and exhibit no colo...

Page 19: ...s a high signal the LED turns on for the duration that TTL IN is held high 3 3 4 TTL OUT BNC Connector Outputs a triggering signal for external software When the LED is off TTL OUT is high when the LE...

Page 20: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 18 This page intentionally blank...

Page 21: ...L5 system is required to prevent excessive dust accumulations Wipe all exterior surfaces with a dry soft cotton cloth Periodically inspect all cables ensuring that all connections are made well and co...

Page 22: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 20 This page intentionally blank...

Page 23: ...only at the factory The cost of shipment both ways is paid for by Sutter Instrument during the first three months this warranty is in effect after which the cost is the responsibility of the customer...

Page 24: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 22 This page intentionally blank...

Page 25: ...utter Information for wavelength availability and other information 4 4 Microscope Adapters Liquid light guide adaptors for various microscopes are available Please contact Sutter Instrument for more...

Page 26: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 24 This page intentionally blank...

Page 27: ...above the fuse holder and below the power receptacle and pry the holder straight out of the power entry module Figure 4 1 Power entry module and fuse location on the rear panel of the Lambda HPX L5 c...

Page 28: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 26 This page intentionally blank...

Page 29: ...put 400 700 nm white light Lamp life 25 000 hours at 80 original output Input Output TTL and analog Dimming PWM or Current Control LED Wavelength Cool white Optional LEDs 630 530 460 and 405 nm Electr...

Page 30: ...0mm Time Delay fuse EIC 60127 2 E g Bussmann GDC S506 2 R or S506 2A RoHS Other Dimensions Controller 12 25 x 9 x 5 25 in 311 x 229 x 133 mm Head attaches to scope 2 5 x 3 in 64 x 76 x 127 mm Length o...

Page 31: ...ains 28 fuses replacement mains 3 I indicators 15 status display 15 input analog 27 TTL 27 voltage 28 installation 9 unpacking 9 Installation 9 instructions 9 introduction 7 description 7 features 7 t...

Page 32: ...entry module 25 power switch 16 precautions 3 electrical 3 PWM frequency 27 S safety warnings 3 electrical 3 safety warnings precautions operational 4 safety warnings precautions 3 4 shuttering 27 st...

Reviews: