LAMBDA DG-4 & DG-5 OPERATION MANUAL – REV.
3.05C
(
20150414
)
iii
DISCLAIMER
The ultra high speed wavelength switching xenon arc lamp Lambda DG-4/5 system is
designed for the specific use as a microscope illuminator and optical wavelength switcher, and
no other use is recommended.
This instrument should only be used in a laboratory environment for use on animal tissues.
It is not intended for, nor should it be used in, human experimentation or applied to humans
in any way. This is not a medical device.
Do not open or attempt to repair the instrument without expressed and explicit instructions
from Sutter Instrument Company. Extreme heat and high voltages are present and could
cause injury.
Do not allow unauthorized and or untrained operatives to use this device.
Any misuse will be the sole responsibility of the user/owner and Sutter Instruments assumes
no implied or inferred liability for direct or consequential damages from this instrument if it
is operated or used in any way other than for which it is designed.
SAFETY WARNINGS AND PRECAUTIONS
Electrical
Operate the Lambda DG-4/5 using 110-120 V AC, 60 Hz, or 220-240 V AC., 50 Hz line
voltage. This instrument is designed for connection to a standard laboratory power outlet
(Overvoltage Category II), and because it is a microprocessor--controlled device, it should
be accorded the same system wiring precautions as any 'computer type' system. A surge
protector and power regulator are recommended.
Fuse Replacement: Replace only with the same type and rating:
5 x 20 mm glass tube, time delay (medium), time lag, or “slow blow”)
Rating:
Line Voltage: 100 - 120VAC
(Voltage Switch Setting: 110V)
Line Voltage: 220 - 240VAC
(Voltage Switch Setting: 220V)
Manufacturer Examples
Manufacturer Examples
RoHS Compliant (Lead Free)?
RoHS Compliant (Lead Free)?
Xenon
Lamp
Power
Rating
Fuse
Rating
Yes No
Fuse
Rating
Yes No
125
Watts
3.0A,
125V
Bussmann: GMC-3-R
Littelfuse: 239 003.P
Bussmann: GMC-3A
Littelfuse: 239 003
T2.0A,
250V
Bussmann: S506-2-R
Littelfuse: 218 002.P
Bussmann: GDC-2A
Littelfuse: 218 002
175
Watts
4.0A,
125V
Bussmann: GMC-4-R
Littelfuse: 239 004.P
Bussmann: GMC-4A
Littelfuse: 239 004
300
Watts
5.0A,
125V
Bussmann: GMC-5-R
Littelfuse: 239 005.P
Bussmann: GMC-5A
Littelfuse: 239 005
T3.15A,
250V
Bussmann: S506-315-R
Littelfuse: 218 3.15P
Bussmann: GDC-3.15A
Littelfuse: 218 3.15
A spare fuse is located in the power input module. Please refer to the fuse-replacement
appendix for more details on fuse ratings and for instructions on how to change the fuse.
1
NOTE: The 125-Watt lamp version of the DG-4/DG-5 is a discontinued product.
Summary of Contents for Lambda DG-4
Page 3: ......
Page 4: ......
Page 9: ......
Page 14: ......
Page 18: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 4 This page intentionally blank ...
Page 26: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 12 This page intentionally blank ...
Page 48: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 34 This page intentionally blank ...
Page 63: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 49 ...
Page 64: ......
Page 70: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 56 This page intentionally blank ...
Page 72: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 58 This page intentionally blank ...
Page 82: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 68 This page intentionally blank ...
Page 86: ...LAMBDA DG 4 DG 5 OPERATION MANUAL REV 3 05C 20150414 72 NOTES ...