16
Once the rod is touching it, externally rotate the cutting mask by 3°.
Note I:
The medial-lateral displacement resulting from this off-
setting must be carried out as deemed appropriate by
the physician, seeking to centre the femoral component,
although it must be noted that rotating towards the ex-
ternal part has the advantage of reducing the tension on
the patella and is conducive to it being centred on the
femoral component.
Note II:
The cutting mask has housings for 4, 8 and 12 mm dis-
tal supplements, which must be installed to provide pro-
per distal support to the cutting mask, mainly in those
cases where recuts have been necessary on the distal
condyles.
3° of external rotation
With the knee at 90° flexion, and the posterior condyles resting on the surface of the tibia which has previously been cut, insert the 3° external rotation
system into the posterior cutting slots of the cutting guide.
Fit the alignment bar into the hole 3° to the left or right. When this bar is aligned with the axis of the tibia, externally rotate the cutting guide by 3°.
This support also has a hole at 0° which may be used at the discretion of the surgeon.
Surgical Technique
Set 3.
Summary of Contents for Genutech CCK
Page 1: ...Surgical Technique Genutech CCK total knee revision system ...
Page 2: ......
Page 36: ...34 FEMUR Insertion of 70 mm stems in femur F 70 F 70 Surgical Technique ...
Page 37: ...35 Insertion of 70 mm stems in the femur greater intramedullary stability F 70 120 200 F 70 ...
Page 38: ...36 Insertion of 120 mm stems in femur F 120 F 120 Surgical Technique ...
Page 55: ......