9
Delayline Detector DLD6060-8S Manual
Delayline Detector DLD6060-8S Manual | Surface Concept GmbH
• In most cases the detector head is stabilized again vibrations by a PTFE inlet within the transport
container to protect the detector head during transportation. In some cases the detector head might
still be stabilized by three transportation locks in the lower aluminum ring.
Remove the three M3 screws transportation locks (where applicable) before mounting
the Delayline Detector (
). Otherwise the analyzer can be seriously damaged.
Figure 1: Removal of the transportation locks.
• When looking into the opened detector flange of the spectrometer from the bottom side, you can
locate the “alignment hole” for the detector. Align the pin in the cover of the Delayline Detector with
the corresponding hole in the ground plate (see
• Be sure that the detector is nearly parallel to the ground plate and push the detector carefully into its
seat.
• While pressing the detector into its seat, make sure that the distance between the two flanges is approx.
4 mm before you feel that the springs are being compressed.
• Hold the detector in its position and tighten the flanges with the M8 screws together.
Figure 2: Orientation pin in DLD cover (for illustration only) and corresponding hole in the ground plate of a PHOIBOS Analyzer.