SLC-TRQ4 manual v1.1                                                                                                                                 

5

 

 

 

SETUP – A LIITLE MORE IN THE DETAILS 

Measuring modes: 

 

 

 

Continuous – records all measuring points 
from start to stop 
Result:  maximum value.  
This  mode  is  ideal  for  monitoring  longer 
procedures  with  several  peaks  as  chuck 

torque testing  

Single  peak  –  Measurement  stops 
automatic after the 1

st

 peak.  

Result: maximum value 
This  mode  is  ideal  for  monitoring  short  
procedures  with  only  1  peak  as  cap 

release torque testing

 

 

 

Double peaks – Measurement stops 
automatic after the 2

nd

 peak. 

Results: Max1, Max2 
This mode is ideal for monitoring short  

Multiple  peaks  –  Records  only  the 
maximum values of the test cycles – 1 test 

cycle  is  represented  by  1  point  on  the 
diagram.  This  test  is  generally  used  for 

Summary of Contents for Smart Load Cell TRQ4

Page 1: ...SLC TRQ4 manual v1 1 1 SMARTLOADCELL USER S MANUAL Smart Load Cell TRQ4 manual desktop torque tester Serial numbers TRQ4 torque tester ST E Date of purchase Warranty 3 years...

Page 2: ...the left side The green LED with start blinking once it has turned on TRQ4 torque testers have built in LCD screens They display the current and the peak values Actual and Maximum To change the setup...

Page 3: ...g unit trigger high and low limits sample rate and turn off time and then the device is ready to be used without the smartphone as well Touchscreen functions Press the Conti Peak button to change betw...

Page 4: ...p the diagram area to start recording 4 Perform your test Continuous and multiple tests can be stopped by tapping the diagram area again Single and Double peak tests will stop automatically 5 After th...

Page 5: ...Single peak Measurement stops automatic after the 1st peak Result maximum value This mode is ideal for monitoring short procedures with only 1 peak as cap release torque testing Double peaks Measurem...

Page 6: ...r the diagram s scaling default setting is the maximum of the cell recommended value is a little above your expected maximum 5 Select sample rate 6 25 400 Hz recommended 50 200 Hz 6 Set turn off time...

Page 7: ...ttery is enough for about 10 hours of operation Please note that when battery is almost discharged it is possible that you experience some spikes in the measurement It is recommended not to discharge...

Page 8: ...thin the tolerance range 1 calibration is ok If not load cell must be recalibrated Recalibration with loads 1 Make sure all the calibration accessories are present and the SmartLoadCell is in the cali...

Page 9: ...green to blue SmartLoadCell application freezes Enter settings apps in on your Android device and force stop the application If this doesn t help it is recommended to uninstall the application and th...

Page 10: ...SLC TRQ4 manual v1 1 10 Fax 36 1 3917310 mail sureotrque eu www suretorque eu...

Reviews: