Use the
TouchRing
or
PW
knob to change the priority on the
overlapped images between live ultrasound and the registered slice of the
3D volume from the 2nd modality.
7#(-
1. Click on
One Point
to better match the live ultrasound with the 2nd
modality 3D volume.
2. Find an anatomical point on the live ultrasound (left side)
3. Click on
Ultrasound
under Record Position to freeze the ultrasound
image
4. Click on the anatomical point to be registered on the ultrasound image
5. Move the probe to find the 2nd modality slice with the same
anatomical point
6. Click on
2nd Modality
under Record Position to freeze the image
7. Click on the anatomical point to be registered on the 2nd modality
The live Ultrasound image is now perfectly synchronized with the 2nd
Modality on a specific anatomical point.
When you want to register more than one internal marker, go to the
Internal Markers Registration section.
Use the
TouchRing
or
PW
to change the priority between the
ultrasound image and the 2nd modality (Fusion blending control).
"7
After registration and particularly before using the system for the current
session, it is mandatory to check the system registration accuracy by
performing a side-by-side comparison of the ultrasound live image and
the registered 2nd modality image using the TouchRing or PW Overlap
controls, to determine how well the two images match.
If the ultrasound live image and the corresponding 2nd modality image
appear to be poorly registered, repeat the registration procedure.
@'-A
C;
Summary of Contents for AixplorerUltimate
Page 1: ... ...
Page 2: ......
Page 3: ... 3 B 3 9 9 9 B 3 B B 0 1 1 B 2 1 3 4 9 1 3 4 C 3 5 6 0 B 5 6 0 9 0 1 A ...
Page 5: ... 7 C 1 CB 8 BD 7 9 B 7 4 B B B B3 B C C C BB D 7 D 7 7 4 7 D 9 D3 7 B 5 7 B D D DB A ...
Page 6: ... A ...
Page 7: ... A ...
Page 8: ... A ...
Page 14: ...C A ...
Page 15: ... A ...
Page 16: ... A ...
Page 23: ... A ...
Page 24: ... C A ...
Page 27: ...General Purpose Sensor omniTRAX CT omniTRAX MR A ...
Page 38: ... A ...
Page 39: ... A ...
Page 40: ... A ...
Page 44: ...Then place the transmitter clamp on the plate as follows C A ...
Page 45: ...Make sure the clamp is well mounted on the plate as shown below A ...
Page 57: ...3 A B ...
Page 58: ...B A ...
Page 101: ... A B ...
Page 102: ... 9 A ...
Page 107: ... A ...
Page 108: ... A ...
Page 113: ... A ...
Page 114: ... C A ...
Page 118: ......