7
* The temperature difference between thermocouple
probe and thermometer will cause an inaccurate
measuring result. Therefore, for the best measuring
and accuracy performance, whenever changing a probe
or plugging a new probe, thermal equivalent between
probe plug and meters input socket is a necessary
condition. Thermal equivalent procedure may take few
minutes and apply only when the probe has been exposed
to an ambient temperature different from the meter.
4-3 Humidity & Ambient Temperature
Measurement
1) Power on the instrument by pressing the ON Button
(3-1, Fig. 1).
2) Select the Relative Humidity function by pressing
Function Button ( 3-6, Fig. 1 ).
3) At the mean time the reading value of relative humidity
and temperature will be displayed on the LCD display.
4) When the meter is applied in a new environment, a few
minutes are required to reach a stable condition.
4-4 Light Measurement
1) Power on the instrument by pressing the ON Button
(3-1, Fig.1).
2) Select the Light Measurement function by pressing the
Function Button ( 3-6, Fig. 1 ) until the light value is
displayed. The light display digits are oriented 180° from
the other function displays for easy exposure and output
reading of the light sensor.
3) Press the Lux/Ft-cd Button (3-5, Fig.1) to
select measuring unit Lux or Ft-cd .