PowerTalk2 User Manual - E1801_20181018 Rev.A
6 Troubleshooting
When a connection to the WheelDrive fails, follow these steps:
1.
Close PowerTalk
2.
Remove USB cable from WheelDrive or PC.
3.
Restart the WheelDrive
4.
Reconnect the USB cable
5.
Restart PowerTalk
6.
If “USB” or “Live” icons below are not green, press connect button
If the steps described above do not work, the failed connection can be related to the COM port number that is used.
PowerTalk only accepts COM port numbers from 1 to 9.
Follow the instructions in this hyperlink to change the COM port number on the PC:
https://thinksmartbox.com/answer/changing-the-com-port-for-a-usb-device/
Note:
The comport numbers will increase every time a new WheelDrive is connected, so it is advised to start with
number 1.