Model 7621

Instruction  Manual

 

 

Summary of Contents for 7621

Page 1: ...Model 7621 Instruction Manual 1 ...

Page 2: ...are utilized to insure that we provide our cus tomers with the most accurate and thoroughly tested product which can be produced As with any new product if is important that you read the instructions com pletely and familiarize yourself with the product s operation before using it If you experience any difficulty in the use of this monitor please call our toll free customer service help line 1 800...

Page 3: ... Parts 6 Notes for Proper Measurements 8 Description of Display Marks 10 Preparation for Measurements 12 Measurements 14 16 Exhaust Velocity 18 More about Blood Pressure Measurements 20 22 24 P r e c a u t i o n s 2 6 Specifications 26 ...

Page 4: ...Identification of Parts ...

Page 5: ...uations of your blood pressure during the course of a day The blood pressure is usually lowest upon arising in the morning before beginning daily activities It will gradually fall again during the sleeping hours 6 Although such cases are rare for subjects having an extremely weak pulse or irregular pulse errors may result which prevent proper meas urement If such abnormal variations are noticed co...

Page 6: ...rement was be gun but pressure was insufficient OAppears when the bat tery voltage is ex cessively low 0 Appears w h e n the blood pressure could not be measured ac curately Appears when the pulse could not ac c u r a t e l y b e m e a s u r e d 0 Flashing when exhaust speed is above 8 mm Hgls I i I j Corrective action Measurement in progress remain quiet P r e s s t h e e x h a u s t b u t t o n ...

Page 7: ...othly and snugly Place the cuff so that the inflatable bag is centered over the brachial artery which is located approximately midway on the arm in the bend of the elbow when the palm is facing upward The bottom edge of the cuff should be positioned from 1 to 2 inches above the elbow Fasten the cuff securely Rolling up a shirt or jacket sleeve to a l l o w t h e a r m c u f f t o b e f i t t e d m...

Page 8: ...octor s office A beeper will sound when pressure reaches 160 200 240 and 280 mmHg 4 When pressurization is completed the automatic exhaust mechanism will gradually reduce arm cuff pressure and the mark will appear on the display indicating that measurement Pressure value appearson is in progress the left and exhaust speed on the right side o ft h e display i Current Pressure xhaus t Speed g r a d ...

Page 9: ...in the figure use a small pm or nee dle insert into tiny hole on top and flip off screw cover then proceed to Step 1 1 Apply the arm Cuff to the left arm and pressurize to the pressure 6 0 mmHg above your suspected systolic pressure Hold the area in and insert a screwdriver into groove 2 If the exhaust velocity is too fast justing screw one fraction of a turn to the left and if it is too slow turn...

Page 10: ...etc prior to measurement can affect the results 4 The blood pressure varies constantly throughout the day Measurement should be made regularly at the same time each day 5 Do not be too impressed by the results of one measurement Keep a record of blood pressure variations Many readings tell a story 6 When making repeated measurement the arm becomes heavy with blood resulting in wide variations in m...

Page 11: ...e varies constantly and may change according to different activities Your blood pressure may change suddenly due to phys ical or mental anxieties smoking exercise eating stress or a number of other conditions Only a qualified Physician can determine whether your blood pressure is normal Hypertension or blood pressure which is most common among adults and the old if left unattended can cause health...

Page 12: ...pressures vary greatly both on a daily and a seasonal basis These variations are even more pronounced in hypertensepa tients Normally the blood pressure rises while at work and is at its lowest during the sleeping period The graph below illustrates the variations in blood pressure over a whole day with measurement taken every five minutes 160 r 140 E 120 f 100 I 80 60 i 40 20 PM Time AM Shown is d...

Page 13: ...M o d e l 7 6 2 1 T y p e Oscillometric Display Digital Measurement range 20 280mmHg Pressure 40 200 pulses minute Pulse Accuracy i3mmHg or 2 whichever greater Pressure 5 Pulse Pressurization M a n u a l pressurizing Depressurization Constant Air Release Valve System Deflation Manual push button Power sources Type AAA batteries 4 Battery life Approx 6 months with 3 min usage per day operating envi...

Page 14: ...anada Return prepaid to Solaray Division of Sunbeam Corporation Canada Limited 266 Grand River Avenue Brantford Ontario N3T 4Y4 Made in Japan CUSTOMER SERVICE HELP L I N E We want to be sure you are completely satisfied with your new pur chase If you haveany questions of experience any problems please call our toll free help line In U S A only phone 1 800 621 8854 between 8 00 a m 12 p m and 1 00 ...