should already be in place and your instrument is ready to use. If you
purchased an adapter separately, simply screw it into the threaded hole
at the opposite end from the light Switch. Your instrument is ready to
use.
4 Operation
In this manual, we will explain general use of the Fiber Microscope. However,
it is important to read the specific termination instructions published by the
manufacturer of the connector you are using. This will ensure proper
termination. Also be sure you have read the User Safety Warning and fully
understand that direct eye contact with laser light and magnified laser light
should be avoided.
1) Begin by loading the polished connector into the adapter.
If you have purchased our Universal Adapter, simply slide the connector
ferrule (2.5mm) into the hole until you feel it hit a stop.
Note:
It may require a little pressure to get it started as our Universal adapter is
equipped with a no-slip gripping mechanism; it is important to slide the connector
ferrule all the way to the end. This will minimize any refocusing that needs to be done
when changing from one connector to the next.
If you have purchased one of the dedicated adapters, insert the
connector just as you would into a patch bay or mated connector.
2) Now turn the light on. The light works by pressing and holding the light
Switch in. You can tell whether the light is on by checking the "Backlight
Window". If the light passes through this window, you are ready to work.
3) Once the connector is loaded and the light is on, hold the eyepiece up to
your eye and begin focusing. Unlike many instruments of this type, the
Fiber Microscope is always centered so there is no alignment to be
done.