SA5052 Digital Multimeter User ‟s Guide
Shijiazhuang Suin Instruments Co., Ltd. 55
Preset and configure the temperature measurement and this command wil not enable the
mesurement For this measurement, resolution is fixed as 3 1/2 digits.
CONFigure?
Query the current configuration and return a quoted charater string.
4.3.2 FUNCtion commands
FUNCtion"<function>”
Select measurement function. The function content among the command strings need to be
indicated by quotation mark, like (FUNC “VOLTDC”). All functions contents include:
VOLTage:DC
VOLTage:AC
CURRent:DC
CURRent:AC
RESistance ( 2-wrie resistance)
FRESistance (4-wire resistance)
FREQuency
PERiod
CONTinuity
DIODe
TEMPerature
CAP
FUNCtion? Query measurement function and return quoted strings.
<function>:RANGe{<range>|MIN|MAX}
Select range for current measurement. For frequency and period measurement, range is
applicable for input voltage not frequency (please use FREQuency:VOLTage, or
PERiod:VOLTage). MIN is selected as minimum range and MAX is as maximum range.