![Suez Sievers M500e Operation And Maintenance Manual Download Page 89](http://html1.mh-extra.com/html/suez/sievers-m500e/sievers-m500e_operation-and-maintenance-manual_1389721089.webp)
I
NSTALLATION
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
89
SUEZ © 2022
Figure 3-8: Asynchronous Serial Port (RS-232)
Installing an Ethernet Connection (Optional)
Ethernet connectivity is available for:
•
Exporting historical data from the Analyzer to a computer, computer network, or
Programmable Logic Controller (PLC).
•
Printing to most PDF printers.
Attach one end of an Ethernet cable to the Ethernet port located on the left outside
panel of the Analyzer (
Figure 3-7: Input and Output Connections
). Attach the other end
of the cable to an Ethernet port on the network, to a computer, or to a printer. After
installing the Ethernet cable, configure Ethernet in the Analyzer’s firmware,
an Ethernet Connection” on page 157
. To set up the printer for use with the Analyzer,
“Enabling a PDF Printer” on page 100
. To assist in using the Modbus
communications with the Analyzer, refer to
Appendix E, "Modbus with the Analyzer."
To install an Ethernet cable
Provide either a cross-over Ethernet cable (for connecting directly to a computer
or printer) or a straight-through Ethernet cable (for connecting to a network).
1. Attach one end of the Ethernet cable to the
Ethernet
port on the Analyzer. (See
“Input and Output Connections” on page 87
.) Attach the other end of the cable
to an Ethernet port on a computer, a printer, or a network.
2.
Configure Ethernet output later in this installation process during
Configure Basic Analyzer Settings.”
NOTE:
The Analyzer’s USB port is not designed for Ethernet
connectivity (using an Ethernet to USB adapter). For acceptable
uses of the USB port, refer to the section
Connection for Data Transfer (Optional)” on page 88
.