Installing a USB Connection for Data Transfer
(Optional)
The USB port provides data transfer to a USB memory device (for use in transferring
exported data into a spreadsheet or database program) or from a USB memory device
for firmware upgrades. The USB port does not facilitate connectivity to a computer.
To install a USB memory device, insert it into the USB port. (See
.) The device must be oriented properly to slide into the
port.
NOTE:
Due to significant variance among USB memory devices,
not all USB memory devices are compatible with the Analyzer.
Installing a Serial Connection (Optional)
The serial connection is provided as an optional means of capturing data from the
Analyzer. Serial cables are available at many computer and office supply retail stores.
Figure 3-7: Input and Output Connections on page 87
, and
8: Asynchronous Serial Port (RS-232)" on page 89
, as needed.
To install a serial connection
1. Connect the male end of the serial cable to the port on the Analyzer labeled
RS-232
and tighten the captive screws.
2.
Connect the other end (female end) of the cable to the serial port on the
computer and connect the captive screws.
NOTE:
For long distance connections (up to 1,000 feet or
305 meters), SUEZ recommends use of either a RS-232 current
loop converter or an RS-422 converter. Optionally, an Ethernet
connection is available. See
Table 3-4: I/O Panel 9-Pin Male D Connector
Pin Number
Signal
Function
2
RXD
M500 Normally Closed Contact
3
TXD
Data from the M500 Output
5
GND
DC Common
Others
(N/C)
No Connections