Working with 4-20 mA Analog (J4) Outputs
The data from the Analyzer can be recorded using the 4-20 mA outputs located on the
J4 terminal block. Refer to the following pin-out information (
) and typical
circuit applications (
) when installing 4-20 mA Analog output wiring.
Instrument Powered Loop Output
Nominal Operating Range: 4-20 mA
Current Calibration Adjustment Range: 0-24 mA
Alarm Indication (option): 22 mA
Maximum Loop Resistance: 600 Ohms
I
SOLATION
L
EVELS
:
This interface does not provide safety barrier isolation. The
isolation is low voltage and intended to isolate the instrument
DC power from the customer’s power supplies only:
•1000 V for 1 second
•Operation rating: SELV limits < 42.5V peak, or 60 VDC
Table 3-2: 4
-
20 mA Analog Outputs (J4)
Pin Number
Function
1
#1 4-20 mA (Return) -
2
#1 4-20 mA (Out) +
3
#2 4-20 mA (Return) -
4
#2 4-20 mA (Out) +
5
#3 4-20 mA (Return) -
6
#3 4-20 mA (Out) +