22
6.6 PRION Program
A. Before start Sterilization program please refer to “6.2 Prepare Sterilization” section.
B. How to set the PRION program
Figure 16
C. Press
to start “PRION Program”. The corresponding sterilization time will be:
Sterilization Time
18min.
Sterilization Temperature
134°C
Dry Time
15 min.
D. Press the START Button
, the setting temperature, duration, and dry time displayed two
times on the three-digit display, and then the sterilizer will automatically run through the
selected program. The current progress of the sterilization cycle is indicated by the
illuminated LED on the Process Status Indicator. (Figure 17)
Figure 17
On completion, the buzzer will sound same time the display will flash. The program has
finished when the buzzer STOPS and the COMPLETE indicator light is illuminated on the
Process Status Indicator.
Warning:
If the COMPLETE indicator light was not lit, the cycle has failed and should
be run again.
134°C
120°C
103 °C
(Room Temp.)°C
0 bar
(kgf/cm
2
)
Vacuum