Operating Instructions

1.  Remove the Microcentrifuge from the box and examine all of the contents for breakage or miss-

ing parts. Save all packaging and paperwork.

2.  Place the Microcentrifuge on a fl at, stable surface. The unit should be placed in a 

non-hazardous environment and located near an electrical outlet.

3.  With the power switch in the off position, attach the power cord to the back of the 

Microcentrifuge and to the grounded electrical outlet.

4.  Turn the power switch to the "ON" position. The Microcentrifuge will automatically perform a 

safety check on the locking mechanism. Open the lid by pressing the illuminated button located 

on the front of the unit.

5.  Insert your pre-loaded sample tubes into the rotor. Make sure all tubes are fi lled to equal levels 

and the lids are closed. Tubes must be balanced properly.

6.  Close the lid and the Microcentrifuge will begin the centrifugation cycle. IMPORTANT

:

 If 

there are loud / unusual sounds or excessive vibrations discontinue operation immediately. 

See the Troubleshooting section in this manual. If you cannot correct the problem, contact 

Cole-Parmer for assistance. 

DO NOT CONTINUE TO OPERATE!

7.  When centrifugation is completed, once again, press the illuminated button on the front of unit to 

stop the spin cycle motor and open the lid. Do not open lid until rotor comes to a complete stop! 

Remove your samples from the rotor.

8.  The Microcentrifuge lid is secured by two locking pins that prevents accidental lid opening during 

operation and greater safety for the operator. When the Centrifuge has completed the spin cycle, 

an automatic lid unlock will occur. If the lid does not open, press and hold the button on the front 

of the unit for one full second, the Centrifuge will unlock the lid again. Should the lid fail to open 

by pressing the front button or in the event that the power to the Microcentrifuge is interrupted, 

the Microcentrifuge lid can also be opened manually by inserting a fl at head screw driver into the 

bottom of the unit and moving the release lever to the right. This will release the pins and allow 

the lid to open. See Figure 1, page 6. 

9.  The Microcentrifuge has been designed with an automatic shut-off feature that discontinues 

operation after 60 minutes of continuous use. This will prevent the unit from running 

continuously if left unattended.

For proper operation, samples must always be loaded into the rotors in a balanced fashion. The 

tubes should be filled to the same level, lids closed and placed opposite each other in the rotor. 

Tubes and strips must always be loaded symmetrically. 

Rotor Installation and Replacement 

The Microcentrifuge features snap on rotors that can be installed and removed without an Allen 

wrench. There are no set screws to tighten or loosen. 
1. Make sure the Microcentrifuge's On/Off switch is in the "Off" position.
2.   Line triangular opening on the rotor with the rotor shaft and press down until the rotor snaps in 

place. Pull up on the rotor to remove.

3.   Turn the unit back to the "On" position, load tubes per the operating instructions and close the lid 

to begin centrifugation.

CAUTION: Spin Balanced Loads Only

CAUTION: Do Not Open Lid Until 

Rotor has Completely Stopped

1065Y17M_Stuart_Microcentrifuge_Manual_ENGLISH.indd   5

01/11/2019   3:26:52 PM

Summary of Contents for Microcentrifuge 8500

Page 1: ...Microcentrifuge 8500 Operator s Manual RoHS 1065Y17M_Stuart_Microcentrifuge_Manual_ENGLISH indd 1 01 11 2019 3 26 52 PM...

Page 2: ...t No SCF4 V05 Power Cord 1 US 1 EU 1 UK Operator s Manual 1 each Catalog Numbers 17414 70 SCF4 Maximum speed 8500 rpm Maximum G Force RCF 3884 Capacity 8 place rotor for 0 5 2 0 mL tubes 4 place 8 str...

Page 3: ...se microcentrifuge tubes and tube strips made from plastic designed to withstand centrifuged forces of at least 3884 g s RCF Always only use original rotors power cord tube adapters and any other spar...

Page 4: ...to the equipment rating plate to ensure that the plug and fusing are suitable for the voltage and wattage stated The wires in the power cable are as follows 230V 120V BROWN HOT LIVE BLACK HOT LIVE BL...

Page 5: ...When the Centrifuge has completed the spin cycle an automatic lid unlock will occur If the lid does not open press and hold the button on the front of the unit for one full second the Centrifuge will...

Page 6: ...isopropyl alcohol The housing and lid can be cleaned with a damp cloth and a mild non corrosive detergent After cleaning ensure all parts are dried thoroughly before attempting to operate the unit Tr...

Page 7: ...xceed price of product This includes liability claims of loss of use lost profits or any exemplary or punitive damages Repair Should your Stuart Microcentrifuge require service or repair please contac...

Page 8: ...Class A IEC EN 61326 1 2013 IEC EN 61010 1 2010 Safety requirements for electrical equipment for measurement control and laboratory use Part 1 General requirements Particular requirements for laborato...

Reviews: