Footswitch Functions

PEDALS (factory default settings)

A
Reverse

Depress pedal for variable speed (up to
maximum selected on console).  Hand-
piece operates in the counterclockwise
direction.

B
Forward

Depress pedal for variable speed (up to
maximum selected on console).  Hand-
piece operates in the clockwise direction.

PADS (factory default settings)

The activated handpiece (TPS or Endo) determines
the operation of the pads.

  LEFT PAD

Irrigation (TPS Handpieces)

a.  Tap pad to enable handpiece irrigation for the
activated TPS handpiece. Irrigation flows only
while handpiece is running.  Tap pad again to stop
flow.
b.  Depress pad until console beeps twice.
Irrigation flows immediately prior to handpiece
operation and continues during handpiece
operation.  Tap pad again to stop flow.

Oscillate (Endo Handpieces)

Tap to allow forward or reverse pedals to operate
Endo handpiece in the oscillate mode.

II   

CENTER PAD

Handpiece (Endo/TPS)

Tap pad to activate desired handpiece.

III   

RIGHT PAD

Attachment Select (TPS Handpieces)

Scrolls the console screen through a list of
attachments.   Stop when you reach the attach-
ment that is assembled to the handpiece.

High/Low Speed (Endo Handpieces)

Toggles control between the high and low speed
settings selected on the console.
High SpeedHandpiece operates up to high
speed selected on console.
Low Speed Handpiece operates up to low
speed selected on console.

Summary of Contents for TPS 5100-8

Page 1: ...9001 269 323 7700 1 800 253 3210 European Authorized Rep RA QAManager StrykerFrance ZACSatolasGreenPusignan Av deSatolasGreen 69881MEYZIEUCedex France Footswitch REF 5100 8 European Equivalent 5100 8 713 Rev B Japanese Equivalent 5100 8 721 Rev B N European Equivalent 5100 8 732 Rev A ...

Page 2: ...ndswitch can be used to operate TPS handpieces The TPS Footswitch eliminates the need for a circulating assistant to actuate the con sole and allows the surgeon to select system functions from within the sterile field The TPS Footswitch controls functions of the activated handpiece Since both TPS and Endoscopy handpieces can be operated from this system the footswitch has dual function pads with f...

Page 3: ...ain to stop flow b Depress pad until console beeps twice Irrigation flows immediately prior to handpiece operation and continues during handpiece operation Tap pad again to stop flow Oscillate Endo Handpieces Tap to allow forward or reverse pedals to operate Endo handpiece in the oscillate mode II CENTER PAD Handpiece Endo TPS Tap pad to activate desired handpiece III RIGHT PAD Attachment Select T...

Page 4: ...t Make sure cable plugged into console connector is fully seated Footswitch is damaged Return to Stryker for repair See Instrument Repair Loaner Program Outside the USA contact your nearest Subsidiary Specifications Model No 5100 8 Footswitch Electrical Input 5VDC 18mA Output 2 5VDC to 4 9VDC IPX7 Protection against temporary immersion Size 11 7 in 297 mm Wide 9 1 in 231 mm Deep 1 9 in 48 2 mm Hig...

Reviews: