Operation
Starting the pump
WARNING
•
Always fill the pump reservoir with water before operating. Do not plug in the product until you fill the water reservoir.
•
Always plug the product into a hospital grade wall outlet.
•
Do not attach the power cord to any moving parts of the bed frame or any location that may pinch or cause power
cord damage. Improper routing can cause power cord damage and possible electrical shock.
•
Always set the temperature as directed by a physician.
•
Always use the
T/Pump
with
Mul-T-Pad
accessories. Do not use with competitor pads.
CAUTION
•
Federal law (USA) restricts this device to sale by or on the order of a physician.
•
Do not cover the product with blankets, pillows, or other insulating materials. Air flow is required to maintain system
performance.
•
Closely monitor the areas of body that are beneath the thermal transfer devices. Also check for areas of the body
for use of topical solutions, or other heat sources. These may interfere with the treatment prescribed.
To start the pump:
1.
Before filling the pump reservoir, attach a pad to the connector hose (see
Connecting single or multiple pads on
). Make sure that there are no kinks in the hose or pad.
2.
Open the hose clamps.
3.
Open the fill cap on the top of the pump.
4.
To fill for cooling
•
Fill with cold water to the Cooling water line.
•
Fill with ice to the full capacity of the reservoir.
Note:
When using the pump for
cooling
, the ice will eventually melt and the setpoint will start flashing.
5.
To fill for heating, fill with room temperature water to the heating water line.
6.
Plug the pump into a properly grounded hospital grade wall outlet.
7.
Press the
On/Standby
button. The light next to the selected temperature begins to flash.
8.
Use the keypad to set the temperature as directed by the physician.
Notes
•
After setting the temperature, press and hold the Setpoint lock button for two seconds to lock the setpoint.
•
If you go past the desired setpoint, keep pressing the button to start at the beginning of the setpoint values.
•
It takes about 15 minutes to reach the selected water temperature. The light next to the selected temperature
becomes steady.
•
Check the water level. If the water level drops below the operating level, add water.
9.
Apply the
Mul-T-Pad
to the patient as directed by a physician.
10. Position the pump at or above the level of the pad.
Note:
Do not place the pump below the pads, water will drain into the pump during shut off. If the pump has been
overfilled or if you connected multiple pads, excess water can leak.
Summary of Contents for T/Pump Series
Page 2: ...sample text ...
Page 4: ...sample text ...
Page 12: ...English EN 1 8 101515 REV E www stryker com Introduction Serial number location ...
Page 28: ...sample text ...
Page 30: ...sample text ...
Page 38: ...Français FR 2 8 101515 REV E www stryker com Introduction Emplacement du numéro de série ...
Page 53: ......