To Drain Water From Handpiece
Remove the battery pack and cutting accessory from
the handpiece.
Using a brush with stiff, non-metallic bristles and
hospital enzymatic cleaner, scrub the debris from the
handpiece. Pay special attention to crevices and hard
to reach areas such as seams, joints, and details
around the blade retainer, trigger, and connector
areas.
Rinse all external surfaces of the handpiece under
running water. Hold the handpiece upright to prevent
water from running into the battery contact area.
If water leaks into the handpiece, tip the handpiece
back as shown to allow drainage from a small open-
ing in the battery contact area.
1.
2.
3.
4.
To Clean Handpiece
Visually inspect the handpiece for any remaining
debris; if any debris is present, repeat the cleaning
and rinsing procedure using fresh hospital enzymatic
cleaner.
Dry the handpiece with a lint-free towel.
After cleaning, sterilize as directed. See
Sterilization
Recommendations
.
5.
6.
7.
To Clean Battery Packs and Accessories
See the care instructions supplied with the battery
packs, battery pack modules and battery charger.
Cleaning Recommendations
WARNINGS:
• Clean and sterilize handpieces, and batteries before
first and every use.
• Prior to cleaning and sterilization, remove all acces-
sories from the handpieces.
• DO NOT use solvents, lubricants, or other chemicals
unless otherwise specified.
CAUTIONS:
• DO NOT immerse a handpiece or battery pack in
liquid. Moisture may enter the equipment, cause cor-
rosion, and damage the electrical and/or mechanical
components.
• DO NOT allow liquid to run directly into any electrical
connection. Moisture may cause corrosion to electri-
cal and/or mechanical components.