6
www.stryker.com
EN
6205-001-700 Rev-D
3. Once properly aligned, insert the attachment into
the handpiece until the attachment “snaps” into
place.
4. Tug the attachment to ensure it is secure.
5. Install the desired cutting accessory. See the
instructions for use manual for the attachment.
To Install Battery Pack
NOTES:
▪
This handpiece is designed to accept a Stryker
Large or Small Battery Pack only (REF 6215-000-
000 or REF 6212-000-000). These battery packs
can be charged in the Stryker System 6 Battery
Charger (REF 6110-120-000) or the Universal
Battery Charger (REF 7110-120-000) configured
with the appropriate battery charger modules.
▪
See the instructions for use supplied with the
battery charger and/or battery pack for charging
details and specifications.
1. Slide a fully charged battery pack firmly into the
handpiece until the battery latch snaps, indicating
the battery pack is secure (see figure 2).
Figure 2 – To Install Battery Pack
Instructions
To Install Attachment
WARNING:
To prevent inadvertent
running of the handpiece, ALWAYS place
the F/R/Safe control in the safe mode
position before installing or removing any
attachment.
NOTE:
Several attachments are available and each
has a specialized accessory retainer.
1. Slide the F/R/Safe control to the safe mode
position.
2. To install an attachment, align the driveshaft of
the attachment with the handpiece spindle. If the
attachment has notches, also align the notches
with the tabs in the sleeve of the handpiece (see
figure 1).
Driveshaft
Tabs
Notches
(not always
present)
Figure 1 – To Install Attachment