To expedite returns, always include the following information with the returned equipment:
•
Contact name, address, phone number
•
Repair purchase order number
•
Part and serial number(s)
•
Detailed reason for return
T
Trro
ou
ub
blle
essh
ho
oo
ottiin
ng
g
W
WA
AR
RN
NIIN
NG
G -- Do not disassemble, modify, service, or repair any equipment without the
authorization of Stryker. Failure to comply may result in electric shock or fire. For assistance,
contact Stryker.
T
Ta
ab
blle
e 1
1 –– C
Crra
ad
dlle
e T
Trro
ou
ub
blle
essh
ho
oo
ottiin
ng
g
O
Ob
bsse
errvva
attiio
on
n
C
Ca
au
usse
e
C
Co
orrrre
eccttiivve
e A
Accttiio
on
n
Erratic system behavior while
equipment is in use.
Electrical noise and/or radio
interference and/or high
frequency (HF) surgical
equipment is present.
Reorient or relocate electrical
equipment to maximize the
spatial distance between the
equipment. Shielding the
location may also be
necessary.
Connect the system and
other operating room
equipment to different
hospital-grade electrical
power receptacles on
different branch circuits.
Turn off all electrical
equipment not in use in the
operating room.
Significant reduction in
battery run time of system.
System batteries are at end
of life.
Contact Stryker.
Cradle is not charging.
Power cord connection issue.
Verify the power cord is
properly connected to the
hospital-grade electrical
power receptacle.
Electrical power not present.
Follow local healthcare
facility protocol and verify
electrical power is present at
the hospital-grade electrical
power receptacle.
Cradle is damaged.
Contact Stryker.
Tablet is not charging.
Tablet is not fully docked in
the cradle.
Verify the tablet is properly
docked, see
EN
20
700001138880 Rev-AC