Reprocessed Ultrasonic Focus™ Curved Shears 

 

5. 

The instrument operates with either the foot switch or hand control.  To operate using hand control, select the 
hand activation button on the generator.  The hand activation button will be illuminated on the generator for the 
hand control buttons to be active.  For further detailed and operation instruction, refer to the Generator 300 
(GEN04) Operator’s Manual.  

6. 

The stand-by button on the generator should be depressed to remove the generator from stand-by. 

7. 

Clean the instrument blade and clamp arm throughout the procedure by activating the tip in  a  sterile saline 
solution for optimum performance.  

a. 

Note

:  Take care not to touch the activated instrument to metal. 

b. 

Note:  

The instrument may be wiped with a sterile moist gauze sponge to remove tissue but do not clean   

the instrument with abrasives. 

8. 

Place the generator in Standby mode to remove visible tissue in the clamp arm, use hemostats to remove the 
tissue.  Depress the standby button on the generator to return the instrument into use. 

9. 

Depress one of the foot pedals or one of the hand control buttons to ultrasonically energize the instrument blade. 

10. 

The selected minimum power level is activated by pressing the left foot pedal of the footswitch or the proximal 
hand control button (MIN).  The selected maximum power level is activated by pressing  the right foot pedal of the 
footswitch or distal hand control button (MAX). 

11.  Note: 

Premature blade failure could occur from scratches on the blade. 

  

       a. When instrument is in use, avoid accidental contact with other instruments.

 

       b. Use only the Torque Wrench to attach or detach the instrument from the hand piece.

 

12. 

Close the clamp arm by closing the finger rings together and insert the shaft through the incision. 

13. 

The  Reprocessed  Ultrasonic  Curved Shears  can be used for dissection, grasping, coagulation, and cutting 
between the blade and clamp arm.  

14. 

For backcutting, use the top of the blade. 

a. 

Note

:  When backcutting, or while the blade is active without tissue between the blade and tissue pad, 

keep the clamp arm open to avoid damage to the tissue pad. 

 

Disassembly 

1. 

At the power switch turn the generator OFF or enter the Standby mode.

 

a.  Close the clamp arm and place the Torque Wrench over the distal end of the shears. 
b. 

Note: 

Take care to avoid injury from the blade tip while placing the Torque Wrench onto or off of the 

shears. 

2. 

Attach  the  Blue Grip Assist on the Hand Piece  and  loosen the instrument by turning the Torque Wrench 
counterclockwise.  Continue to loosen by turning the instrument manually to completely remove the instrument 
from the Hand Piece.  

3. 

Remove the Torque Wrench. 

4. 

Dispose of the Torque Wrench, Grip Assist and the instrument in an appropriate container.  

 
Storage and Handling 

 

Store at room temperature. 

 

Avoid prolonged exposure to elevated temperatures. 

 

Summary of Contents for Reprocessed Ultrasonic Focus

Page 1: ...ral U S A law restricts this device to sale by or on the order of a physician STERILE Explanation of Icons Sterilized by Ethylene Oxide Gas Date of Reprocessing Use by Date Product Code Do Not Reuse See Instructions For Use Stryker Sustainability Solutions Inc 2006 2011 Phoenix AZ 85044 sustainability stryker com 888 888 3433 ...

Page 2: ... hazards consult the medical literature Prior to the initiation of the procedure verify compatibility of minimally invasive instruments and accessories when they will be used together since there may be variations from manufacturer to manufacturer The use of these instruments requires a thorough understanding of the techniques and principles of laser electrosurgical and ultrasonic procedures Inapp...

Page 3: ...ect the instrument for overall condition and physical integrity Do not use the instrument if any damage is noted Return the instrument and packaging to Stryker Sustainability Solutions if it is not in acceptable condition for surgery 4 Attach the Blue Grip Assist on the Hand Piece to the instrument by rotating the instrument onto the Hand Piece in a clock wise rotation as viewed from the distal en...

Page 4: ... power level is activated by pressing the right foot pedal of the footswitch or distal hand control button MAX 11 Note Premature blade failure could occur from scratches on the blade a When instrument is in use avoid accidental contact with other instruments b Use only the Torque Wrench to attach or detach the instrument from the hand piece 12 Close the clamp arm by closing the finger rings togeth...

Page 5: ...al numbers or other identification marks have been removed or destroyed If Stryker determines in its reasonable discretion that the claimed defect or non conformance in the product is excluded from warranty coverage as described hereunder it will notify the customer of such determination TO THE FULLEST EXTENT PERMITTED BY LAW THE EXPRESS WARRANTY SET FORTH HEREIN IS THE ONLY WARRANTY APPLICABLE TO...

Reviews: