Appendix A: EMC Information
Radiated RF EM
Fields
IEC61000-4-3
10 V/m 80 MHz to 2,7 GHz
80 % AM at 1 kHz
385-6000 MHz, 9-28V/m, 80%
AM(1kHz) pulse mode and
other modulation
10V/m
Recommended separation distance
d=
√
P 150kHz to 80MHz
d=0.6
√
P 80MHz to 800MHz
d=1.2
√
P 800 MHz to 2.7G Hz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
b
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey ,
a
should be
less than the compliance level in each
frequency ranged.
Interference may occur in the vicinity
of equipment marked with the following
symbol:
NOTE 1
: U
T
is the a.c. mains voltage prior to the application of the test level
NOTE 2
: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 3
: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people
NOTE 4
: The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795
MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. The amateur radio
bandsbetween 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz, 5.3 MHz to 5.4 MHz, 7 MHz to
7.3 MHz,10.1 MHz to 10.15 MHz, 14 MHz to 14.2 MHz, 18.07 MHz to 18.17 MHz, 21.0 MHz to 21.4 MHz, 24.89 MHz to
24.99MHz, 28.0 MHz to 29.7 MHz and 50.0 MHz to 54.0 MHz.
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the device.
b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Summary of Contents for P100
Page 2: ......
Page 21: ...www stryker com 476004 5210 V1 5 21 Return To Table of Contents ...
Page 24: ......
Page 43: ...www stryker com 476004 5210 V1 5 21 Retour à la Table des matières ...
Page 46: ......
Page 65: ...Retur til Indholdsfortegnelse www stryker com 476004 5210 V1 5 21 ...
Page 68: ......
Page 87: ...www stryker com 476004 5210 V1 5 21 Zurück zum Inhaltsverzeichnis ...
Page 90: ......
Page 109: ...www stryker com 476004 5210 V1 5 21 Volver al Índice ...
Page 112: ......
Page 131: ...www stryker com 476004 5210 V1 5 21 Palaa Sisällysluettelo ...
Page 134: ......
Page 153: ...www stryker com 476004 5210 V1 5 21 Torna a Indice ...
Page 156: ......
Page 175: ...www stryker com 476004 5210 V1 5 21 Terug naar de Inhoudsopgave ...
Page 178: ......
Page 197: ...www stryker com 476004 5210 V1 5 21 Tilbake til Innhold ...
Page 200: ......
Page 219: ...www stryker com 476004 5210 V1 5 21 Regressar ao Índice ...
Page 222: ......
Page 241: ...www stryker com 476004 5210 V1 5 21 Powrót do Spis treści ...
Page 244: ......
Page 263: ...www stryker com 476004 5210 V1 5 21 Tillbaka till Innehållsförteckning ...