![Stryker nGenius Universal Tracker Instructions For Use Manual Download Page 5](http://html.mh-extra.com/html/stryker/ngenius-universal-tracker/ngenius-universal-tracker_instructions-for-use-manual_1385977005.webp)
www.stryker.com
Specifi cations*
Model:
REF 6000-014-000 nGenius Universal Tracker
Size:
74.2 mm [2.92 in] length, 46.2 mm [1.82 in] width, 36.1 mm [1.42 in] height
Weight:
49 g [1.73 oz] without Instrument Battery
Material:
Titanium (electronics housing), PPSU (battery holder, front cover, SELECT button), stainless
steel (interface release button)
Power supply:
3V ---
,
internally powered (lithium battery)
Enclosure protection:
IPX0 Ordinary Equipment
LED classifi cation:
Class1, IEC 60825-1 Edition 1.2 2001-08, Safety of Laser Products Part 1:
Equipment
Classifi cation Requirements and User’s Guide
Operation
Storage and
Transportation
RF-ID operating frequency:
125 kHz
4
Only use the manual cleaning method.
Refer to the
Guide for Cleaning and Steam-Based Sterilization
REF 6000-005-750 for cleaning safety and cau-
tion notes, cleaning equipment as well as detailed
cleaning and inspection instructions.
For the nGenius Universal Tracker follow the
procedure for Manual Cleaning and Disinfection
of Instruments with Electronics as described in
section 5.
The nGenius Universal Tracker is a Cleaning
Group IV instrument as defi ned in the
Guide for
Cleaning and Steam-Based Sterilization
.
Sterilization Instructions*
Only use the Sterrad® Sterilizer. Follow the
Sterrad® protocol instructions.
Do not steam sterilize.
-20°C
75°C
* Specifi cations listed are approximate and may vary slightly from unit to unit.
Cleaning and Inspection Instructions