Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 6 of 61
Equipment
Specification or Description
Manufacturer or Part number or
Catalog number
Defibrillator analyzer with external noninvasive
pacer measurements*
Energy range: 0 to 450 J
Load resistance: 50
ohms±1% Accuracy:
+/- 2%, non-
inductive
Waveforms: Simultaneous 12-lead output
Rates: 30 bpm, 120 bpm, with rate
accuracy of ± 1% Amplitude: 1 mV ±
5%, based on Lead II
ECG performance: Amplitudes of Lead II and
Leads V1-V6 are equivalent. Lead I = 70%
amplitude of Lead II.
Sine wave: 10 Hz @ 1 mV ± 2%, based on
Lead II
Fluke® Biomedical Impulse
7000DP with QUIK-COMBO
adapter accessory 16/7 D/P
ADPT104**
Safety Analyzer
90 V ac rms to 264 V ac rms mains voltage
Current range: 0-1999
A
Current accuracy: 5% of reading or 1 digit
(whichever is greater)
Fluke Biomedical ESA612*
Decade resistance box
0 to 9 M
resistance box
Resolution: 1
; accuracy: ± 1%
IET RS-200 Resistance Substitute*
Remote Sync Test Pulse Generator
***
21300-007574
SPO2 Leakage Cable
21300-007561, 21300-005487,
21300-007562
AC Power Cable
Commercial
Accessory
–Test Plug, Quik-Combo
11113-000002
Paddle Assembly - Detachable, LP20
11130-000037
QUIK-COMBO Therapy Cable
11110-000040
50 mm Recorder Paper
11240-000013