LIFEPAK
®
20
Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 56 of 61
PIP-Direct Applied Part Leakage Test-SPO2
Note:
Execute this test if the LIFEPAK 20 is equipped with SpO2
WARNING
POSSIBLE EQUIPMENT DAMAGE
Do not defibrillate when connected to the safety analyzer!
1. Turn the device ON.
2. Press the LEFT/RIGHT arrows to select the RL lead, and set the Safety Analyzer controls as follows:
3. Press the TEST button to measure the Direct Applied Part Leakage current.
4. Verify the device AC Mains LED is ON.
5. Verify the measured current is less than 2625
A (120 VAC and 240 VAC).