Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 54 of 61
PIP-Direct Applied Part Leakage Test-ECG
WARNING
POSSIBLE EQUIPMENT DAMAGE
Do not defibrillate when connected to the safety analyzer!
1. Turn the device ON.
2. Press the LEFT/RIGHT arrows to select the RA lead, and set the Safety Analyzer controls as follows:
3. Press the TEST button to measure the Direct Applied Part Leakage current.
4. Verify the device AC Mains LED is ON.
5. Verify the measured current is between 2
A and 45
A (120 and 240 VAC).