LIFEPAK
®
20
Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 51 of 61
PIP-Direct Equipment Leakage Test - Single Fault Condition (SFC)
WARNING
POSSIBLE EQUIPMENT DAMAGE
Do not defibrillate when connected to the safety analyzer!
1. Turn the device ON.
2. Press the
A button on the Safety Analyzer.
3. Press the F1/DIRECT EQUIPMENT button on the Safety Analyzer and set the Safety Analyzer controls as follows:
Note:
Pause briefly between switching polarity to prevent damage to the defibrillator.
4. Verify the device AC Mains LED is ON.
5. Verify the measured current is between 15
A and 270
A (120 VAC) or between 15
A and 450
A (240 VAC).