Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 22 of 61
Note:
Perform this test if 5-lead ECG cable is used. Otherwise, skip to the next test.
The 5-Lead ECG tests consist of:
PIP - 5-Lead ECG Leads Off Detection Test
PIP - 5-Lead ECG Gain Test
PIP - 5-Lead ECG Leads Off Detection Test
1.
Connect the 5-wire ECG cable between the device and Impulse 7000DP as
shown in Figure 1.12.
2.
Set the Impulse 7000DP output to a 1-mv, 10-Hz sine wave.
3.
Set the device
Lead
selection to
LEAD II
.
4.
Remove the RL lead from the Impulse 7000DP, and verify the device displays
an
ECG LEADS OFF
message and a repeating priority 3 tones shall sound
when the Lead is removed. Reconnect the RL lead.
5.
Remove the RA lead from the Impulse 7000DP, and verify the device displays
an
RA LEADS OFF
message and a repeating priority 3 tones shall sound when
the Lead is removed. Reconnect the RA lead.
6.
Remove the LL lead from the Impulse 7000DP, and verify the device displays
an
LL LEADS OFF
message and a repeating priority 3 tones shall sound when
the Lead is removed. Reconnect the LL lead.
7.
Set the device lead selection to
LEAD I
.
8.
Remove the LA lead from the Impulse 7000DP.
9.
Verify the device displays an
LA LEADS OFF
message and a repeating priority
3 tones shall sound when the Lead is removed. Reconnect the LA lead.
10.
Set the device
LEAD
selection to
C LEAD
.
11.
Remove the C1/V1 Lead from the Impulse 7000DP.
12.
Verify the device displays a
C LEADS OFF
message and a repeating priority 3
tones shall sound when the Lead is removed. Reconnect the C1/V1 lead.
13.
Continue to the next test with this setup in place.
Figure 1.12: 5 Lead ECG Leads Test
Setup