PIP - Keypad Verification Tests
To perform Keypad Verification Tests:
1. Select
TESTS
in the Service menu.
2. Select
BUTTONS
in the
Service/Tests
submenu as shown in Figure 1.6.
3. Press each front panel button when prompted by the flashing button
legend (although you may press the buttons in any order).
4. Verify with each button pressed that its associated text box is highlighted.
Note:
A failure is indicated by a text box that is not highlighted. It is
normal for the buttons with up/down arrows to highlight only the
arrows.
5. Press the switch located between the
OPTIONS
and
PAUSE
buttons.
Note:
The switch is hidden in the elastomer keypad.
6. Verify the
TEST COMPLETE
message appears on the bottom of the
screen and the
Service LED
is not on.
7. Press the
SPEED DIAL
to exit at the end of the test.
8. Continue with the next test while still in Service mode.